COM Server Commands
241
SCPI.SENSe(
Ch
).FREQuency.CENTer
Object Type
Property (read/write)
Data Type
Double
Target
Channel
Ch
,
Ch
Description
The stimulus center value of the sweep range for linear or logarithmic
sweep types.
Range
From the minimum value to the maximum value of the analyzer
frequency range.
Out of Range
Sets the value of the limit, which is closer to the specified value.
Preset Value
The center value of the analyzer frequency range.
Unit
Hz (Hertz)
Syntax
Dim Value As Double
Value = app.SCPI.SENSe(Ch).FREQuency.CENTer
app.SCPI.SENSe(Ch).FREQuency.CENTer = 1Е9
Equivalent
Softkeys
Stimulus > Center