FSG/MTM/001 060511
99
4.2.5
Check and Adjustment of Scan-Width
Images of the Scan-Width adjustments are shown in:
Fig. 4-20 COUGAR 25, CHAMELEON 25, Scan-Width, page 102
Fig. 4-21 COUGAR 36, CHAMELEON 36, PREMIER 36, Scan-Width (A), page
103
Fig. 4-22 COUGAR 36, CHAMELEON 36, PREMIER 36, Scan-Width (B), page
103
Fig. 4-23 CHROMA Tx 40, CRYSTAL Tx 40, Scan-Width (A), page 104
Fig. 4-24 CHROMA Tx 40, CRYSTAL Tx 40, Scan-Width (B), page 104
Fig. 4-25 CHROMA Tx 40, CRYSTAL Tx 40, Scan-Width (C), page 105
Fig. 4-26 MAGNUM XL 54, Scan-Width (A), page 106
Fig. 4-27 MAGNUM XL 54, Scan-Width (B), page 106
Fig. 4-28 MAGNUM XL 54, Scan-Width (C), page 107
Fig. 4-29 MAGNUM XL 54, Scan-Width (D), page 107
Fig. 4-30 COUGAR Tx 36, CHAMELEON Tx 36, PREMIER Tx 36, Scan-Width
(A), page 108
Fig. 4-31 COUGAR Tx 36, CHAMELEON Tx 36, PREMIER Tx 36, Scan-Width
(B), page 108
The aim of this adjustment is to obtain a camera Scan-Width of:
COUGAR 25, CHAMELEON 25:
635 mm
COUGAR 36, CHAMELEON 36, PREMIER 36:
516 mm
CHROMA Tx 40, CRYSTAL Tx 40:
344.4/350.2/344.4 mm
MAGNUM XL 54:
350.9/358.9/358.9/350.9 mm
COUGAR Tx 36, CHAMELEON Tx 36, PREMIER Tx 36: 466.7 mm
The Scan-Width is correct if the total sum of Scan-Width Units (SWU) at the start and at the
end of the scanline is as shown in the table below (
1 SWU is either 1 Black Line or 1 White
Space of the Scan-Width Pattern (A
)).
Model
Optical Adjustment Pattern
Total sum of
Scan-Width
Units
Tolerance
COUGAR 25, CHAMELEON 25
6779D105, sub-pattern A
6
±0.5 SWU
COUGAR 36, CHAMELEON 36,
PREMIER 36
6779D107, sub-pattern A
6
±0.5 SWU
CHROMA Tx 40, CRYSTAL Tx 40
6779D102, sub-pattern A
6
±0.5 SWU
MAGNUM XL 54
6779D130, , sub-pattern A
6
±0.5 SWU
COUGAR Tx 36, CHAMELEON Tx
36, PREMIER Tx 36
6779D127, sub-pattern A
6
±0.5 SWU
The tolerance corresponds to approximately 0.10% on the Scan-Width (scaling error).
Adjustment Procedure
Always do the Focus adjustment before adjusting the Scan-Width, see “Interaction” below.
Summary of Contents for CHAMELEON 25
Page 69: ...FSG MTM 001 060511 65 Fig 3 21 OIBA Component Layout Fig 3 20 OSB Component Layout ...
Page 102: ...FSG MTM 001 060511 98 Fig 4 19 COUGAR Tx 36 CHAMELEON Tx 36 PREMIER Tx 36 Focus Signals ...
Page 106: ...FSG MTM 001 060511 102 Fig 4 20 COUGAR 25 CHAMELEON 25 Scan Width ...
Page 109: ...FSG MTM 001 060511 105 Fig 4 25 CHROMA Tx 40 CRYSTAL Tx 40 Scan Width Camera C ...
Page 118: ...FSG MTM 001 060511 114 Fig 4 32 COUGAR 25 CHAMELEON 25 CCD Centering and Vertical Positioning ...
Page 183: ...FSG MTM 001 060511 179 Fig 11 1 COUGAR 25 Primary Wiring WHITE LME WHITE SMPS L N WHITE ...
Page 185: ...FSG MTM 001 060511 181 Fig 11 3 CHAMELEON 25 Primary Wiring WHITE LME WHITE SMPS L N WHITE ...
Page 187: ...FSG MTM 001 060511 183 Fig 11 5 COUGAR 36 Primary Wiring WHITE SMPS WHITE LME L N WHITE ...
Page 193: ...FSG MTM 001 060511 189 Fig 11 11 MAGNUM XL 54 Primary Wiring ...
Page 194: ...FSG MTM 001 060511 190 Fig 11 12 MAGNUM XL 54 Secondary Wiring ...
Page 205: ...FSG MTM 002 060511 7 Fig 2 1 COUGAR Sx 25 and CHAMELEON Sx 25 Block Diagram ...
Page 207: ...FSG MTM 002 060511 9 Fig 2 2 CHROMA XL 42 and CRYSTAL XL 42 Block Diagram ...
Page 241: ...FSG MTM 002 060511 43 Fig 6 1 Sx 25 Primary Wiring ...
Page 242: ...FSG MTM 002 060511 44 Fig 6 2 Sx 25 Secondary Wiring ...
Page 243: ...FSG MTM 002 060511 45 Fig 6 3 XL 42 Primary Wiring ...