12.  Press Key “4" once  to remove the prompt.   

*The LCD will read “000" (Waiting for operator to key in function number to go to).  

NOTE: If the operator wishes to check the alarms, leave NO2 calibration gas 
flowing into the sensor, otherwise turn off the flow of NO2 calibration gas and inject 
Zero Air into the sensor for 1 minute to purge the sensor of NO2 before proceeding. 
  

13.  Press Key “3" once to enter function number “001" (Exit CONFIG mode). The LCD 
will read “001"
.  

14.  Press Key “4" twice to NO2nfirm function 1 (Exit CONFIG mode).  

15.  The P2621-NO2/VC returns to “Gas Monitoring Mode” with the LCD displaying the 
current    NO2 NO2ncentration.  

VIII) SENSOR REPLACEMENT  

The P2057 NO2 sensor should be replaced when the displayed readings on the P2621-
NO2/VC become erratic or the unit can no longer be calibrated.  Conspec recommends 
replacing all P2057 NO2 sensors in service for 5 years as their continued operation is no 
longer assured. Use the supplied hex-key to loosen the tamper-proof screw in the lower 
left-hand corner of the P2621-NO2/VC’s door.  Use a standard flat blade screwdriver to 
loosen the remaining three door retaining screws.  Open the door of the P2621-NO2/VC 
and unplug the P2057's  4-wire cable from the “J4” Connector on the P2065 Smart 
Monitor Card.  Clip the wire tie securing the cable to the lower left-hand corner of the 
P2065 Smart Monitor Card and discard.  Grasp the ring protruding from the bottom of the 
P2621-NO2/VC and twist counter clockwise (as viewed from below) while holding the 
body of the P2057 inside the enclosure.  Unscrew the outer ring (Note: the outer ring 
holds the dust screen in place inside the diffusion tube.  When the outer ring is removed 
care should be taken to retain the dust screen if the P2057 is to be returned for sensor 
replacement) and remove the P2057 NO2 sensor from the P2621-NO2/VC enclosure.  
Installation of a new P2057 NO2 sensor is accomplished in the reverse order.   
NOTE: When a new P2057 NO2 sensor is installed allow 10 minutes for the sensor 
to warm up.  

IX) TROUBLE SHOOTING  

1. No Lights On The Front Panel

 

Turn OFF the power to the P2621-NO2/VC and to the fan Control system. Open the 
enclosure and check that the 24 VAC power supply wiring is securely attached to the 
terminals T1 and T2, and that the wiring from the Connector terminals to the circuit 
board is securely attached. Turn ON the 24 VAC power to the P2621-NO2/VC.  If the 
power light is OFF, using a voltmeter, measure the voltage between T1 and T2. The 
voltage should read close to 24 VAC. If it does not, the problem is in the power supply 
and not in the P2621-NO2/VC.  If there is 24 VAC across the T1 and T2 terminals and 

Summary of Contents for P2621-NO2/VC

Page 1: ...Conspec Controls Inc P2621 NO2 VC Nitrogen Dioxide Ventilation Fan Controller Operations Manual Conspec Controls Inc 1 800 487 8450...

Page 2: ...the relay coil is energized whenever the P2621 NO2 VC is in a non alarm condition and the normally open contact terminal connections K1 K2 is held closed When any fault occurs the Fan relay is de ene...

Page 3: ...ystem design and installation Zoned systems are easily created when several sets of fans need to be controlled independently 1 Be sure to turn OFF power to fan Control system and 24 VAC power supply b...

Page 4: ...t a regular interval Conspec stipulates a gas injection into the sensor and verify the response of the sensor for every THREE MONTHS of operation To perform an actual gas verification simply inject a...

Page 5: ...g three door retaining screws Open the door of the P2621 NO2 VC and locate the 20 header pin Connector marked J2 in the middle of the left edge of the P2065 Smart Monitor Card Plug the P2621 P program...

Page 6: ...sensor Press Key 4 once to accept the ZERO GAS calibration If the reading on LCD is outside the range of 25 100 the monitor should be sent back to the factory for in house calibration The message SCAL...

Page 7: ...ector on the P2065 Smart Monitor Card Clip the wire tie securing the cable to the lower left hand corner of the P2065 Smart Monitor Card and discard Grasp the ring protruding from the bottom of the P2...

Page 8: ...should be closed low or zero resistance If the circuit is not closed return the P2621 NO2 VC to your supplier or CONSPEC for repair or replacement If the above resistance checks between terminals K1 a...

Page 9: ...A Current Source Control Outputs Optional Strobe TB1 V Strobe P2065 J3 OUT 4 First Alarm Piezo TB1 V Piezo P2065 J3 OUT 3 Second Alarm NOTE The Control Outputs are open collector transistors limited t...

Reviews: