background image

  

5

kit (P1879-NO2/VC) from Conspec and follow the instructions to perform a calibration. 
The user may also send the unit to the distributor or directly to the factory for calibration.  

VII) CALIBRATION PROCEDURE  

NOTE:  All Conspec Smart Gas Monitors are fully tested and calibrated before 

shipment. Calibration is the process of setting two distinct points for the Nitrogen 
Dioxide monitor to reference.  All NO2 readings are derived from these two 
reference points.  The first reference point is referred to as “ZERO GAS” 
representing a total absence of NO2 and is obtained by injecting zero air into the 
sensor.  The seNO2nd reference point is referred to as “SCALED GAS” and can 
be either half-scaled (½ of full range or 10ppm) or full scaled (full range or 
20ppm) NO2 and is obtained by injecting a known concentration of NO2 into the 
sensor. The proper application of the calibration gases is essential to the correct 
operation of the P2621-NO2/VC.

  

Place the strap of the P1879-NO2 Calibration Kit over the P2621-NO2/VC so that 
the case hangs below the monitor.  Use the supplied hex-key to loosen the tamper-
proof screw in the lower left-hand corner of the P2621-NO2/VC’s door.  Use a 
standard flat blade screwdriver to loosen the remaining three door retaining 
screws.  Open the door of the P2621-NO2/VC and locate the 20 header-pin 
Connector marked “J2" in the middle of the left edge of the P2065 Smart Monitor 
Card.  Plug the P2621-P programmer into the “J2" Connector on the P2065 card 
so that the “J2" graphic on the P2621-P plug is located uppermost.  Open the top 
half-door of the P1879-NO2 calibration kit and locate the P3013 calibration 
adapter.  Insert the P3013 calibration adapter into the bottom of the P2057 NO2 
sensor by pushing upward while twisting.  Install a 0.5 liter per minute regulator 
on each gas cylinder.  Connect a 3' section of flexible tubing onto the barb of each 
regulator.  Follow instructions under “Manual Calibration.”   

MANUAL CALIBRATION (This is Function “011" in CONFIGure Mode).  

NOTE: DO NOT ENTER FUNCTION 11 UNLESS A GAS CALIBRATION IS TO BE 
PERFORMED.  STEPPING THROUGH FUNCTION 11 WITHOUT APPLYING 
THE CALIBRATION GASES WILL ERASE THE PREVIOUS CALIBRATION 
SETTINGS RENDERING THE MONITOR INOPERATIVE UNTIL A PROPER 
CALIBRATION IS PERFORMED. 
 

1.   Depress the “CONFIG/4" Key for 2 seNO2nds.  The message “ENTER ACCESS 

CODE” will scroll across the 3 digit LCD .Press Key “4" once to remove the 
prompt. The LCD will read “000"  (Waiting for operator to key in access CODE).  

2.  Press Key 3 once. The LCD will read “001"  (Access CODE factory set.)  

3.  Press Key “4" twice to enter the Access CODE of “001".The message “CONSPEC 

Ver#” will scroll across the LCD (Software Version Number). 

Summary of Contents for P2621-NO2/VC

Page 1: ...Conspec Controls Inc P2621 NO2 VC Nitrogen Dioxide Ventilation Fan Controller Operations Manual Conspec Controls Inc 1 800 487 8450...

Page 2: ...the relay coil is energized whenever the P2621 NO2 VC is in a non alarm condition and the normally open contact terminal connections K1 K2 is held closed When any fault occurs the Fan relay is de ene...

Page 3: ...ystem design and installation Zoned systems are easily created when several sets of fans need to be controlled independently 1 Be sure to turn OFF power to fan Control system and 24 VAC power supply b...

Page 4: ...t a regular interval Conspec stipulates a gas injection into the sensor and verify the response of the sensor for every THREE MONTHS of operation To perform an actual gas verification simply inject a...

Page 5: ...g three door retaining screws Open the door of the P2621 NO2 VC and locate the 20 header pin Connector marked J2 in the middle of the left edge of the P2065 Smart Monitor Card Plug the P2621 P program...

Page 6: ...sensor Press Key 4 once to accept the ZERO GAS calibration If the reading on LCD is outside the range of 25 100 the monitor should be sent back to the factory for in house calibration The message SCAL...

Page 7: ...ector on the P2065 Smart Monitor Card Clip the wire tie securing the cable to the lower left hand corner of the P2065 Smart Monitor Card and discard Grasp the ring protruding from the bottom of the P2...

Page 8: ...should be closed low or zero resistance If the circuit is not closed return the P2621 NO2 VC to your supplier or CONSPEC for repair or replacement If the above resistance checks between terminals K1 a...

Page 9: ...A Current Source Control Outputs Optional Strobe TB1 V Strobe P2065 J3 OUT 4 First Alarm Piezo TB1 V Piezo P2065 J3 OUT 3 Second Alarm NOTE The Control Outputs are open collector transistors limited t...

Reviews: