EPIQ 5 & 7 Training Manual
© 2016 Conquest Imaging
50
If any one of the performance tests fails, perform the
System
Test
. You gain access to the system test using the
Minimum Test
Environment
MTE. The MTE has many different utilities with
different levels of access.
To ensure proper system operation, be sure to turn on all
peripherals before checking system functionality.
Test Equipment
You will need:
FSE Tool Kit
ATS model 539 Phantom
Test Procedures
The initial setup procedure consists of observing system power-up
and initialization and adjusting the monitor for optimal viewing.
These tests are also performed during preventative maintence.
Exam-Start Tests
If patient data security is disabled, you are able to start a regular
exam. If patient data security is enabled, you can only start an exam
as an emergency study using a temporary patient ID.
Successful execution of the exam-start tests assumes the following:
Transducer ports are operational.
Transducers, patient data entry, and the retractable keyboard are
operational.
Touch screen has been successfully calibrated and is operational.
Data security is functioning normally.
To Start an Exam when Data Security is Enabled:
1. Connect and initialize any transducer. If possible, connect
transducers to all four ports.
2. Acquire an image.
3. Touch
Patient
and then click
Use Temporary ID
.
The ATS model 539
Phantom provides ameans
of evaluating imaging
systems.
It has line targets for
distance measurements
and target structures that
mimick tissue structures of
varying sizes and
contrasts.
All manuals and user guides at all-guides.com