EPIQ 5 & 7 Training Manual
© 2016 Conquest Imaging
55
The appropriate licensed options are installed.
To Test Live 3D operation:
1.
Select a 3D transducer and an appropriate application/preset.
2.
Position the transducer over an area of interest.
3.
Press 3D to start the Live 3D acquisition
4.
To select a Live 3D display format, touch a format in Layout.
5.
Optimize image size, position, and rotation, using the trackball
and touch screen controls.
6.
To select a 3D display format, touch a format in Layout.
7.
To start Full Volume acquisition, touch Full Volume.
8.
Press Rotation Style and do one of the following:
a.
Select Absolute to allow rotating the image relative to
the acquisition axes of the image.
b.
Select Relative to allow rotating the image relative to
the display screen.
9.
To reset the orientation of the image views to their original
orientations, touch 3D Home.
10.
Press 3D Opt to select it and turn the know to select the
appropriate setting.
Testing Connectivity and Review
Successful execution of this test indicates:
End Exam and Review are working.
The patient database is functional.
DVD drive is operational.
Transfer of JPEG loops to the DVD and back means that the PCI-E
board is functional.