EPIQ 5 & 7 Training Manual
© 2016 Conquest Imaging
58
To Perform a 2D Distance Measurement:
1. Start a patient exam, if you have not already done so.
2. Obtain a 2D image and press Freeze.
3. Press or touch Measure, and then touch Tools.
4. Touch Distance.
5. Use the trackball to position the caliper for the first end point and
click to anchor it.
6. Use the trackball to position the caliper for the second end point.
The results update as the distance between the calipers changes.
7. To end the measurement, press the middle trackball button.
8. Touch Report to display the patient report.
9. Press 2D to return to 2D imaging.
10. Press Pointer a couple of times. Verify that an arrow appears and
disappears from the display.
Query Retrieval Test
You must be able to complete a query before you can download
exams to an ultrasound system.
To Query a PACS Device for Exams:
1. Verify that query retrieval is properly configured.
2. Touch
Review
to open the
Patient Directory
.
3. On the Source menu, select the device that you want to query.
4. Set the parameters.
Note: The left, right, and
middle trackball buttons
can each be set to one of
three measurement
operations: End, Toggle,
or Next Measure. The
middle trackball button
can also be set to Not
Assigned. To se
t these functions, see the
Help.