EPIQ 5 & 7 Training Manual
© 2016 Conquest Imaging
54
marker.
b.
If Box Size is amber-colored, rotate the trackball and
press the left trackball button to change the color box
size.
c.
If Box Position is amber-colored, rotate the trackball
and press the left trackball button to move the color
box around the display.
3.
Press PW to switch between the 2D/CPA image and the
scrolling Doppler display.
4.
Touch
Simult
. Verify that the 2D/CPA and scrolling Doppler
displays are live.
5.
Turn PW to change the Doppler gain. Verify that Doppler noise
on the display and that you hear Doppler audio.
6.
Touch
Mute
. Verify that there is no sound from the speakers.
Touch
Mute
again to turn the volume on again.
7.
Touch the center dot in the volume control slider and slide
both directions. Verify that the volume changes appropriately.
8.
Verify that your system and transducer support CW Doppler,
and if so, press PW to turn off PW imaging, and touch CW to
start CW Doppler. Verify the proper display.
9.
Turn CW to change the Doppler gain.
10.
Press 2D to return to 2D imaging.
11.
Touch M-Mode to obtain a scrolling M-mode display.
12.
Turn M-Mode to change the M-mode gain.
Testing Live 3D
Successful execution of this test indicates:
The Motor Control Module has been successfully installed and
flashed (if the 3D transducer has a motor).