EPIQ 5 & 7 Training Manual
© 2016 Conquest Imaging
42
Under
Edit DICOM Device
you can enter changes to the
Device Type
Device Name
AE Title
Port No.
IP Address
Click
Advanced Settings
and if necessary, select or change the
required settings.
Confirm the settings and click
Save
.
If you want to perform DICOM verification now, click
Automatically Perform DICOM Verification After Saving?
And
click
Save
again.
If you did not verify the server in the previous step, click
Verify
.
To Configure DICOM Parameters
For systems that are busy, slow, or for older networks setting these
DICOM parameters can help the performance until the network is
updated. In general, only someone who is well versed in DICOM
parameters would alter these. To access these settings:
Press Support and Click the
Network/DICOM
DICOM Settings
.
Select a DICOM server, and then click
Edit
.
Under
Edit DICOM Device
complete any of the following fields:
Max PDU Size
specifies the maximum PDU (protocol data unit)
size.
ARTIM Timeout
specifies the Association
Request/Reject/Release Timer (ARTIM) time-out value.
Network Reply Timeout
specifies the amount of time that the
system will wait for a reply from the server.
Idle Association Timeout
specifies the amount of time
allowed for the system to maintain an idle association with
the server.
All manuals and user guides at all-guides.com