14
Result
TIMER
HIPOT
ON
ON
Test Timer and the breakdown detect are On. Test will stop automatically if the internal
test time counter is completed, or a leakage failure, or breakdown failure.
DEFEAT
ON
Test Timer is defeated. After full voltage is reached, the test will continue only as long as
the Test button is held in, minimum one second. Test will stop automatically on all leakage
or breakdown failures.
ON
DEFEAT
Buzzer will turn sound, It will not allow to test until the Test Timer switch is in defeat.
DEFEAT
DEFEAT
Test will continue only as long as the TEST button is held in, minimum one second. The
HT-5000P ac/dc 200mA will NOT shut down on a dielectric failure, but the front panel
Hipot Fail light will flash to indicate a dielectric breakdown. The Hipot Pass light will not
light at the completion of a successful test. For safety reasons, excessive leakage current,
as set by the Leakage knob, will cause the HT-5000P ac/dc 200mA to shut down.
Table 3-3: Front Panel Switch Truth Table