background image

 

 

ITEM 

NO. 

NAME 

FUNCTION 

HIPOT 

When  ON, Dielectric breakdown detect will be activated. 
When in DEFEAT position, test will continue regardless of a dielectric breakdown, but it can be stopped by the excess 
leakage limit. NOTE: The TIMER switch must also be in the DEFEAT position, otherwise the buzzer will sound. The 
test will continue only as long as the TEST button is pressed.  Minimum test time is approximately one second.   

TIMER 

When ON, test duration is set by TIMER ADJUST, Item 4. 
When in DEFEAT position, testing continues only as long at TEST button is pressed.  Minimum test time is one second.  
NOTE:  TIMER Switch position must be DEFEAT when HIPOT Switch is in the DEFEAT position. 

LEAKAGE ADJUST 

Adjusts the shutdown point for the Leakage Current Test.  For details see “Leakage Current Adjust.”   

TIMER ADJUST 

Adjusts the test duration. For details see “Test Time Adjust”.   

RESET Button 

When lit, indicates that the HT-5000P ac/dc 200mA is unarmed. When the RESET button is pressed, the TEST switch is 
lit.  PRESSING THE RESET BUTTON AT ANY TIME IMMEDIATELY STOPS TESTING. 

TEST Button 

When lit, indicates the HT-5000P ac/dc 200mA is ready to test; press to begin testing. NOTE: The Voltage Adjust knob 
must be at the minimum position in order to disarm the Zero Voltage Start safety feature, Item 15. 

VOLTAGE METER 

Connected to the output. Reads actual output voltage. Adjust meter range with VOLTAGE ADJUST Knob, Item 15. 

HIPOT PASS LED 

Indicates test conclusion with satisfactory results. 

FULL VOLTAGE LED 

Lights when output voltage has ramped up. Test time starts when this indicator lights. 

10 

HIPOT FAIL LED 

Lights when arcing or insulation flashover has occurred.  

11 

EXCESS LEAKAGE LED 

Actual leakage current has exceeded the shutdown point set with Shutdown Limit Potentiometer, Item 3. 

12 

CURRENT METER 

Connected  to  the  output.    Read  the  current  flowing  through  the  return  lead  of  the  HT-5000P  ac/dc  200mA  during  the 
test. 

13 

HIGH VOLTAGE output 
jack 

Connect high voltage lead here to conduct an AC & DC test. 

14 

RETURN Receptacle 

At chassis ground reference level.  Connect black return lead here. 

15 

VOLTAGE ADJUST Knob 

Voltage  is  continuously  adjustable  before  or  during  testing  with  this  knob.  NOTE:  Before  the  test  can  commence,  the 
Zero Voltage Start safety feature most be disarmed by turning the voltage knob to the minimum.  

16 

AC POWER Switch 

Energizes the HT-5000P ac/dc 200mA. 

17 

AC/DC Switch 

 

Selects AC or DC output. Changing switch setting does not change output until the next time the Reset button is pushed. 

 

Table 3-2.  Front Panel Control, Indicators and Connectors 

Summary of Contents for HT-5000P

Page 1: ......

Page 2: ...i...

Page 3: ...nd ergonomic front panel and represents the latest in high voltage laboratory testing To fully appreciate all the features of your new meter we suggest that you take a few moments to review this manua...

Page 4: ...iii...

Page 5: ...ory Testing 10 Factory Settings 10 Voltage Type Selection 10 Display of Leakage Limit and Duration Settings 10 Leakage Current Level Adjust 10 Voltage Adjust 11 High Voltage Test Time Adjust 11 Settin...

Page 6: ...1...

Page 7: ...ge current level pass fail requirement at this time for most equipment However higher than normal leakage current on a particular sample may indicate an assembly and or a component problem in the circ...

Page 8: ...dicator lights a breakdown arc has been detected High Voltage Discharge The HT 5000P ac dc 200mA has an internal circuit designed to remove the high voltage after completion of the dielectric withstan...

Page 9: ...eature most be disarmed by turning the voltage knob to the minimum also the unit must be armed by pressing the red RESET button The test will not begin until the blue TEST Button is pushed If a failur...

Page 10: ...Panel Test Time Leakage Limit Voltage Adjustment Timer ON DEFEAT Hipot ON DEFEAT ENVIRONMENTAL Operating Temperature 15 40 C Relative Humidity Range 0 90 non condensing GENERAL Input power requiremen...

Page 11: ...Table 3 1 Shipment Cables Connectors Use the original shipping container for subsequent shipping If the original shipping container is not available be sure that adequate protection is provided to pre...

Page 12: ...7 Figure 3 1 Front Panel Controls Indicators and Connectors...

Page 13: ...djust knob must be at the minimum position in order to disarm the Zero Voltage Start safety feature Item 15 7 VOLTAGE METER Connected to the output Reads actual output voltage Adjust meter range with...

Page 14: ...current condition 12 Push the TEST button 13 Use the Voltage Adjust knob to increase the voltage output but the test should terminate immediately and the buzzer should sound The Excess Leakage Indica...

Page 15: ...e Selection Use the AC DC Voltage Switch This switch is functional only when the RESET or TEST button is lit Display of Leakage Limit and Duration Settings To view the Test Duration and Leakage Limit...

Page 16: ...nd Consult the safety agencies on the test time for the type of equipment being tested If a different test time is required use this procedure to set it 1 Connect the tester to a correctly rated sourc...

Page 17: ...nt under test is grounded by the HT 5000P ac dc 200mA a risk of shock exists Testing Products This section describes how to conduct a test Testing can be terminated at any time by pressing the RESET b...

Page 18: ...g as the TEST button is pressed 12 If no breakdown is detected the high voltage will be removed at the end of the test the Hipot Pass indicator will light and the RESET button will light 13 Do not dis...

Page 19: ...failures ON DEFEAT Buzzer will turn sound It will not allow to test until the Test Timer switch is in defeat DEFEAT DEFEAT Test will continue only as long as the TEST button is held in minimum one se...

Page 20: ...se is required for all in warranty repairs The manufacturer is also available for calibration and or repair of instruments that are beyond their warranty period Contact the manufacturer for a cost quo...

Page 21: ...was entered the display will read Sure if not start over at step 1 6 While Sure is displayed on the screen you can 6 1 Press Reset to exit out of the Calibration Mode and keep all of the currently pro...

Page 22: ...t 2 Select the type of output voltage with the AC DC Switch 3 Set up a 1000 1 probe with the volt meter has is shown on Figure 4 1 4 Enter to Calibration Mode See Entering Calibration Mode procedure 5...

Page 23: ...firm the new voltage meter calibration performing again the Voltage Meter Verification procedure mentioned before Leakage Meter Verification 1 Turn Off the HT 5000P ac dc 200mA unit 2 Select the type...

Page 24: ...the external current meter mA scale equals the number showed on the front meter of the HT 5000P ac dc 200mA unit then press the Test button and the front display will show hold for a few seconds wait...

Page 25: ...ce West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Reviews: