4
Section 2
Introduction and Specifications
This manual contains complete operating and specifications for the HT-5000P ac/dc
200mA Dielectric Withstand Tester.
The instrument is a bench-type Dielectric Withstand Tester with AC and DC Output,
designed for laboratory testing of components, and insulation systems.
The HT-5000P ac/dc 200mA features automatic two buttons operation, with numerous
safety features designed to protect the operator:
- The test return lead is directly connected to ground potential for operator safety.
- The test can be immediately terminated at any time by pressing the red
RESET
button.
- Before the test can commence, the Zero Voltage Start safety feature most be disarmed by
turning the voltage knob to the minimum, also the unit must be armed by pressing the red
RESET
button. The test will not begin until the blue
TEST
Button is pushed.
- If a failure is encountered, the high voltage is immediately shut down, a buzzer sounds
and the voltage discharge progress is shown by the front panel meter.
- Failure modes are shown by the front panel LED's.
Convenience and testing features include:
- Test time, and leakage limit are settable.
- Test duration timer is defeatable for specialized testing.
- Testing may be terminated or continued when a dielectric breakdown is detected.
- Test results are determined quickly, without operator intervention.
- The HT-5000P ac/dc 200mA allows custom setups for test time and leakage limit.
Your Tester has a warranty for a period of one year upon shipment of the instrument to the