5
Specifications
ELECTRICAL
Voltage Output
AC: 0 - 5000V
DC: 0 - 7000V
Max Leakage Current Output:
AC: 5000V = 200mA
DC: 5000V= 200mA, 7000V=140mA
Leakage Current Limit Adjustment:
AC: 5 – 200 mA
DC: 5 – 200 mA
Pass/Fail Criteria:
Leakage Current:
Pass/Fail point user adjustable.
Dielectric Breakdown:
Separate high frequency detection circuit for
breakdown spike detection
Test Time:
User adjustable 1 to 90 sec. 1 sec steps, defeatable
Voltage Meter Accuracy
AC: ±1.0% FSC from 0.50KV to 5.00KV volts
DC: ±1.0% FSC from 0.50KV to 7.00KV volts
Leakage Meter Accuracy
± 1mA
Duty cycle
100 %
Test adjustments
Front Panel:
Test Time
Leakage Limit
Voltage Adjustment
Timer ON/DEFEAT
Hipot ON/DEFEAT
ENVIRONMENTAL
Operating Temperature
15-40 C
Relative Humidity Range
0-90% non-condensing
GENERAL
Input power requirements
208VAC, 3φ, 50/60Hz, 6A/phase.
Weight
90 lbs.
SAFETY AGENCY TOPICS
Transformer Output
> 500VA
Visual Indication of Voltage Output
Provided by front panel meter, directly connected to
high voltage output
Failure Indication
Audible, provided by internal buzzer
Visual, provided by red LEDs, on front panel
Test can be automatically terminated on failure
Leakage Test
Provided; 5 mA AC factory set pass/fail point, user
adjustable.
Zero Voltage Start
It will NOT let to start the test until the voltage
adjustment knob is at the minimum position.
Table 2-1. HT-5000P ac/dc 200mA Specifications