10
Setting up the HT-5000P ac/dc 200mA for Laboratory Testing
This section describes procedures for setting the HT-5000P ac/dc 200mA tester.
a.
Leakage current level
b.
High voltage level
c.
High voltage test time
d.
Test timer switch
e.
Breakdown detect switch.
This will allow you to change settings from the factory settings below. Refer to table 3-1,
figure 3-1 and table 3-2 for location of items.
Factory Settings
The HT-5000P ac/dc 200mA is configured as shown when shipped from Compliance
West USA:
Voltage Display:
Vrms
Leakage Current Level:
5mA
High Voltage Level:
Minimum
High Voltage Test Time:
1 second
Test Timer switch:
ON
Breakdown Detect switch:
ON
Voltage Type Selection
Use the AC/DC Voltage Switch. This switch is functional only when the RESET or
TEST button is lit.
Display of Leakage Limit and Duration Settings
To view the Test Duration, and Leakage Limit current settings, hold down the
RESET
button for 2
seconds. The meter will display “L” with the Leakage Limit value in mA. Hold down the
RESET
button
again for 2 seconds, and the meter will display “d” with the Test Duration set time in seconds.
Leakage Current Level Adjust
1.
Connect the HT-5000P ac/dc 200mA to a correctly rated source of supply and turn ON the
tester.
2.
Push the
RESET
button. The
TEST
indicator should light, indicating that the HT-5000P
ac/dc 200mA is ready.
3. Turn the
Leakage Limit Adjust
. As soon as the potentiometer starts turning, the voltage
meter will start blinking and displaying “L” with the new value. Leakage Level can be set in
1 mA increments.