Table 3. Panel Controls, Indicators and Connectors. 

ITEM NO.

 

NAME 

FUNCTION 

Appliance  Inlet  /  Fuse  holder    / 
Power Switch  

Use supplied cord set to connect tester to appropriate source of supply.  Replace line fuse.  Turn Tester ON/OFF. 
Power switch must be accessible during operation 

Fuse  replacement  warning  /  Rating 
of supply 

Specifies replacement fuse and required supply voltage. 

Testlink OUTPUT Receptacle 

The Output Testlink Cableset lead is connected here. The cableset is included when you get Testlink option. 

Testlink RETURN Receptacle 

The Return Testlink Cableset lead is connected here. The cableset is included when you get Testlink option. 

Testlink 

Allows connection with GF-30AC-T for Ground Bond Test. A cableset is included when you get this option. 

Interlock 

This is a safety feature, when in short-circuited the HT works regular, when open stops immediately its function. 

Relay Interface 

Relay Interface/PLC Interface. Test signals through relay contacts/coils (Optional).  

Ground Check Switch 

Enables or disables the Ground Continuity test. 
Turn ON for use with three-wire (grounded) power supply cords. 
Turn OFF for use with two-wire power supply cords. 
(When  ON,  conducts  Ground  Continuity  test  between  the  chassis  and  the  grounding  pin  of  the  line  cord  of  the 

equipment being tested.  When OFF, Ground Continuity test is bypassed for testing of double-insulated equipment 
and other types of equipment without a grounding pin in the line cord.) 

Ground Resistance Setting 
Adjustment 

Used to adjust the Ground Continuity pass/fail point setting. See “Ground Continuity” on the Adjustment section for 

more information 

10 

Ground Offset Button 

This  button  can perform  two  different  functions. Pressing  and holding  it  for 1  second,  the  HT-S  will  measure  and 

display on the front display the Ground Continuity resistance of the EUT under test. Pressing and holding the button 
for 5 seconds will compensate the resistance of the test cables of fixture, see “Ground Continuity Offset Adjustment” 
and “Ground Continuity Resistance Measurement” sections for more details. 

11 

Leakage Limit Adjustment 

Used  to  adjust  the  trip  level  for  the  excessive  leakage  current  test.    See  "Leakage  Current  Level"  on  Adjustment 

section for more information. 

12 

Test Time Adjustment 

Used  to  adjust  high  voltage  test  duration.    See  "Adjustment  of  the  High  Voltage  Test  Duration"  section  for  this 

procedure. 

13 

Ramp Time Adjustment 

Used to adjust the amount of time used to increase the high voltage from zero volts to the required level.  See "High 

Voltage Ramp Time" on Adjustment section for correct procedure. 

14 

Directions 

Provides directions for tester operation to test personnel. 

15 

Voltage Adjust 

Used to adjust the High Voltage output.  See "High Voltage Level" on the Adjustment section for correct procedure. 

16 

Calibration resistor 

This resistor (optional) may be used to verify the leakage current. 

 

 
 
 
 

 

Summary of Contents for HT-2000S

Page 1: ...HT S Series Dielectric Withstand Tester With Ground Continuity Check Instruction Manual for Models HT 2000S 0 2000VAC HT 2800S 0 2800VDC HT 3000S 0 2000VAC 0 2800VDC HT 4200S 0 3000VAC 0 4200VDC...

Page 2: ...2...

Page 3: ...g logic controlled circuit design and ergonomic front panel and represents the latest in high voltage production line testing To fully appreciate all the features of your new meter we suggest that you...

Page 4: ...Interface PN 00 RI 19 Interlock PN 00 XI 19 Section 3 20 Operating Instructions 20 Setting up your Tester 20 AC Line Voltage Requirements 20 Fuse Replacement 21 Initial Checkout Procedure 21 High Volt...

Page 5: ...5 Technical Assistance 30 Contact Information 30 Section 5 31 Maintenance and Calibration 31 Service Information 31 Cleaning 31 Calibration Information 31...

Page 6: ...ring the test due to low insulation resistance of the insulating material which separates the two conductors and 2 an abrupt dielectric breakdown due to electrical arcing or discharge either through t...

Page 7: ...that a low resistance ground is provided by the building wiring If the HT S is used on a high resistance grounding circuit dangerous high voltages may be present to the operator In addition the power...

Page 8: ...with his right hand This will allow the greatest separation between the operator and the test being conducted The HT S is designed to bleed the high voltage away after the test has concluded In order...

Page 9: ...emoving the Test Return Lead clip from the equipment under test and reconnecting it taking care to make a good connection The unit should then be retested If a failing result is repeated the connectio...

Page 10: ...leakage current is below the limit set by the rear panel adjustment Sporadic Leakage Current Failures If the red Excess Leakage LED light is on the buzzer sounds and the test is terminated the leakage...

Page 11: ...n a problem has been found with the insulation between primary and ground The equipment under test should be examined reworked and successfully tested before being shipped High Voltage Discharge The H...

Page 12: ...buzzer sounds and any voltage stored in the equipment being tested is bled off by a resistor bank in the HT S unit The voltage discharge progress is shown by the front panel meter Failure modes are s...

Page 13: ...ge Limit Voltage Adjust Ground Continuity Defeat Switch Ground Continuity Limit Ground Continuity Offset Tamper proof operation available by removing adjustment shafts ENVIRONMENTAL Operating Temperat...

Page 14: ...utes to become familiar with the use of its controls indicators and connectors The front panel features of the HT S are shown in Figure 1 and Figure 2 then described in Table 2 Figure 1 Front Panel Co...

Page 15: ...15 Figure 2 Front Panel Controls Indicators and Connectors for HT 4200S...

Page 16: ...ground circuit resistance is acceptable the green continuity LED will light and the test will continue If the Ground Check switch is defeated by the operator both LED s will be lit for the duration of...

Page 17: ...T S tester take a few minutes to become familiar with the use of its controls indicators and connectors The rear panel features of the HT S are shown in Figure 3 and described in Table 3 Figure 3 Rear...

Page 18: ...passed for testing of double insulated equipment and other types of equipment without a grounding pin in the line cord 9 Ground Resistance Setting Adjustment Used to adjust the Ground Continuity pass...

Page 19: ...Output Full Voltage 7 Output Excess Leakage 8 Output Hipot Pass 9 Output Hipot Fail 10 Output Any Failure 11 Output COM2 12 Common Table 4 Relay Interface Pin Configuration Interlock PN 00 XI The Saf...

Page 20: ...essories which meet the manufacturer s specifications shall be used Connect this equipment to grounded outlet only The integrity of the test system grounding is dependent on building ground If reshipm...

Page 21: ...s AC and 2800 Volts DC or up to 3000 Volts AC and 4200 Volts DC for HT 4200S generated by the HT S is exposed during this test A risk of shock exists Exercise care when using the HT S High Voltage Per...

Page 22: ...e High Voltage Device socket may be as much as 2000 volts AC or 2800 volts DC or up to 3000 Volts AC and 4200 Volts DC for HT 4200S Proceed with caution 1 Disconnect all leads of the HT S front panel...

Page 23: ...200S HT 3000S HT 2000S Voltage Type AC Leakage Current Level 5 mA High Voltage Ramp Time 1 second High Voltage Level 1200 Volts High Voltage Test Time 2 seconds Ground Continuity Level 1 50 HT 2800S V...

Page 24: ...of supply and turn the power switch to the I or ON position Push the RESET button The yellow TEST indicator should light indicating that the HT S is ready to test e Use the AC DC button to select the...

Page 25: ...reached The test can be terminated at any time by pressing the RESET button 6 Check the position of the Ground Check switch before testing Setting the Ground Check Switch This switch controls whether...

Page 26: ...t in shortcut circuit the green alligator clip end to the Ground Connector Tip connected on step 3 See Figure 6 5 For HT 4200S plug the supplied 18 AWG GL Test Lead Alligator Clip Banana Plug ends bla...

Page 27: ...sistance measurement is more than 1 50 the display will show OPEn Operation Techniques The following paragraphs describe how to operate the HT S tester Before proceeding with testing the HT S tester s...

Page 28: ...t set on the HT S unit the green Ground OK LED will light and the test will continue 5 2 Ramp the voltage at the rate set by the Ramp Time procedure o If leakage current between the AC pins of the pow...

Page 29: ...y cord and the chassis ground of the unit rises above the value set by the Leakage Limit procedure the red Excess Leakage LED will light the buzzer will sound and the test will terminate o If the leak...

Page 30: ...m Compliance West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D S...

Page 31: ...tion and or repair of instruments that are beyond their warranty period Contact the manufacturer for a cost quotation Ship the instrument and your remittance according to the instructions given by the...

Reviews: