Table 2. Front Panel Controls, Indicators and Connectors. 

ITEM 

NAME 

FUNCTION 

RESET Button / Red Indicator  

When  lit,  indicates  that  the  HT-S  tester  is  unarmed.    This  button  must  be  pushed  before  the 

TEST

  button  is  functional. 

When the 

RESET

 button is pressed, the red 

RESET 

indicator goes out and the yellow 

TEST

 indicator is lit.  PRESSING 

THE 

RESET

 BUTTON AT ANY TIME STOPS THE TEST.    

TEST Button / Yellow Indicator 

When lit, indicates that the HT-S tester is ready to test the connected equipment. The yellow 

TEST

 indicator goes out when 

the 

TEST

 button is pressed. 

Voltage Meter 

Provides visual indication to the operator of the actual output voltage of the HT-S tester.  Used to set the test voltage level 
during the setup procedure.   

Ramp Bar Graph LED’s 

Individual  LED’s  light  in  sequence  from  left  to  right  as  the  test  voltage  ramps  from  zero  to  full  voltage.    LED’s  turn  off 
(ramp back down) at end of test. 

Hipot Pass LED 

At the end of the preset high voltage duration time, if no failures are encountered, the green light will light and the test will 
terminate. 

Full Voltage LED 

If full voltage is successfully reached, the 

Full Voltage

 LED will light and the high voltage duration timer starts. 

Ground OK / Ground Open LED's 

Indicates result of Ground Continuity test between grounding pin of the line cord and exposed metal parts of the equipment 
under test.  If the ground path resistance it is higher than the indicated on the ground setting, or if the leads are not properly 
connected,  the  red 

Ground  Open

  LED  will  light,  the  internal  buzzer  will  sound,  and  the  test  will  be  terminated.    If  the 

ground circuit resistance is acceptable, the green continuity LED will light and the test will continue.  If the Ground Check 
switch is defeated by the operator, both LED's will be lit for the duration of the test. 

Excess Leakage LED 

Indicates failure of leakage current test.  If leakage current between the primary circuit and ground is higher than the preset 
value, the red LED will light, the internal buzzer will sound, and the test will be terminated.   

Hipot Fail LED 

Indicates failure of high voltage test.  If arcing or a flashover of the insulation system between primary parts and ground is 
encountered, the red breakdown LED will light, the internal buzzer will sound, and the test will be terminated.   

10 

Return Lead Receptacle 

Grounded banana plug receptacle.  The 18 AWG Test Return Lead provided is connected here. 

11 

High Voltage Device Receptacle 

NEMA 5-15R orange receptacle.  For connection of the equipment under test. See items 11a and 11b for slots description. 

11a 

High Voltage Output 

On  the  NEMA  plug,  the  two  top  vertical  slots  are  tie  together,  the  high  voltage  to  perform  the  Dielectric  Withstand  test 
comes true these two slots.  

11b 

Ground  Continuity  Measurement 
Point 

On  the  NEMA  plug,  the  bottom  round  slot  is  the  ground  continuity  measured  point.  Ground  continuity  test  circuit  is 
measured from this slot (Item 11b) to the Return Lead Receptacle (Item 10).  

12 

High Voltage Test Jack 

Red  High  Voltage  jack.    For  connection  of  high  voltage  test  lead  for  testing  insulating  materials,  etc.  This  High  Voltage 
Jack is standard on models HT4200S and HT-3000S, and optional on models HT-2800S and HT-2000S 

13 

AC/DC Switch 

Selects AC or DC output.  Changing switch setting does not change output until the next time the Reset button is pushed. 
This option is only available on 

HT-3000S & HT-4200S

  model. 

 

 

Summary of Contents for HT-2000S

Page 1: ...HT S Series Dielectric Withstand Tester With Ground Continuity Check Instruction Manual for Models HT 2000S 0 2000VAC HT 2800S 0 2800VDC HT 3000S 0 2000VAC 0 2800VDC HT 4200S 0 3000VAC 0 4200VDC...

Page 2: ...2...

Page 3: ...g logic controlled circuit design and ergonomic front panel and represents the latest in high voltage production line testing To fully appreciate all the features of your new meter we suggest that you...

Page 4: ...Interface PN 00 RI 19 Interlock PN 00 XI 19 Section 3 20 Operating Instructions 20 Setting up your Tester 20 AC Line Voltage Requirements 20 Fuse Replacement 21 Initial Checkout Procedure 21 High Volt...

Page 5: ...5 Technical Assistance 30 Contact Information 30 Section 5 31 Maintenance and Calibration 31 Service Information 31 Cleaning 31 Calibration Information 31...

Page 6: ...ring the test due to low insulation resistance of the insulating material which separates the two conductors and 2 an abrupt dielectric breakdown due to electrical arcing or discharge either through t...

Page 7: ...that a low resistance ground is provided by the building wiring If the HT S is used on a high resistance grounding circuit dangerous high voltages may be present to the operator In addition the power...

Page 8: ...with his right hand This will allow the greatest separation between the operator and the test being conducted The HT S is designed to bleed the high voltage away after the test has concluded In order...

Page 9: ...emoving the Test Return Lead clip from the equipment under test and reconnecting it taking care to make a good connection The unit should then be retested If a failing result is repeated the connectio...

Page 10: ...leakage current is below the limit set by the rear panel adjustment Sporadic Leakage Current Failures If the red Excess Leakage LED light is on the buzzer sounds and the test is terminated the leakage...

Page 11: ...n a problem has been found with the insulation between primary and ground The equipment under test should be examined reworked and successfully tested before being shipped High Voltage Discharge The H...

Page 12: ...buzzer sounds and any voltage stored in the equipment being tested is bled off by a resistor bank in the HT S unit The voltage discharge progress is shown by the front panel meter Failure modes are s...

Page 13: ...ge Limit Voltage Adjust Ground Continuity Defeat Switch Ground Continuity Limit Ground Continuity Offset Tamper proof operation available by removing adjustment shafts ENVIRONMENTAL Operating Temperat...

Page 14: ...utes to become familiar with the use of its controls indicators and connectors The front panel features of the HT S are shown in Figure 1 and Figure 2 then described in Table 2 Figure 1 Front Panel Co...

Page 15: ...15 Figure 2 Front Panel Controls Indicators and Connectors for HT 4200S...

Page 16: ...ground circuit resistance is acceptable the green continuity LED will light and the test will continue If the Ground Check switch is defeated by the operator both LED s will be lit for the duration of...

Page 17: ...T S tester take a few minutes to become familiar with the use of its controls indicators and connectors The rear panel features of the HT S are shown in Figure 3 and described in Table 3 Figure 3 Rear...

Page 18: ...passed for testing of double insulated equipment and other types of equipment without a grounding pin in the line cord 9 Ground Resistance Setting Adjustment Used to adjust the Ground Continuity pass...

Page 19: ...Output Full Voltage 7 Output Excess Leakage 8 Output Hipot Pass 9 Output Hipot Fail 10 Output Any Failure 11 Output COM2 12 Common Table 4 Relay Interface Pin Configuration Interlock PN 00 XI The Saf...

Page 20: ...essories which meet the manufacturer s specifications shall be used Connect this equipment to grounded outlet only The integrity of the test system grounding is dependent on building ground If reshipm...

Page 21: ...s AC and 2800 Volts DC or up to 3000 Volts AC and 4200 Volts DC for HT 4200S generated by the HT S is exposed during this test A risk of shock exists Exercise care when using the HT S High Voltage Per...

Page 22: ...e High Voltage Device socket may be as much as 2000 volts AC or 2800 volts DC or up to 3000 Volts AC and 4200 Volts DC for HT 4200S Proceed with caution 1 Disconnect all leads of the HT S front panel...

Page 23: ...200S HT 3000S HT 2000S Voltage Type AC Leakage Current Level 5 mA High Voltage Ramp Time 1 second High Voltage Level 1200 Volts High Voltage Test Time 2 seconds Ground Continuity Level 1 50 HT 2800S V...

Page 24: ...of supply and turn the power switch to the I or ON position Push the RESET button The yellow TEST indicator should light indicating that the HT S is ready to test e Use the AC DC button to select the...

Page 25: ...reached The test can be terminated at any time by pressing the RESET button 6 Check the position of the Ground Check switch before testing Setting the Ground Check Switch This switch controls whether...

Page 26: ...t in shortcut circuit the green alligator clip end to the Ground Connector Tip connected on step 3 See Figure 6 5 For HT 4200S plug the supplied 18 AWG GL Test Lead Alligator Clip Banana Plug ends bla...

Page 27: ...sistance measurement is more than 1 50 the display will show OPEn Operation Techniques The following paragraphs describe how to operate the HT S tester Before proceeding with testing the HT S tester s...

Page 28: ...t set on the HT S unit the green Ground OK LED will light and the test will continue 5 2 Ramp the voltage at the rate set by the Ramp Time procedure o If leakage current between the AC pins of the pow...

Page 29: ...y cord and the chassis ground of the unit rises above the value set by the Leakage Limit procedure the red Excess Leakage LED will light the buzzer will sound and the test will terminate o If the leak...

Page 30: ...m Compliance West USA is available Phone 800 748 6224 Hours 8 30 AM 4 30 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D S...

Page 31: ...tion and or repair of instruments that are beyond their warranty period Contact the manufacturer for a cost quotation Ship the instrument and your remittance according to the instructions given by the...

Reviews: