46
IE-MLG-M1200E-01
Thermocouple of the “K” type (NiCr-Ni)
Measured quantity:
temperature
Range:
(-
200 to 1,300) °C
Accuracy without probes:
± (0.3% from the measured value + 1.5 °C)
Resolution:
better than 0.1
°C (
to achieve this value, the
corresponding number of decimals must be set
)
Cold junction:
compensated in the temperature range of
(-
10 to 60) °C
Thermocouple of
the “J” type (Fe-Co)
Measured quantity:
temperature
Range:
(-
200 to 750) °C
Accuracy without probes:
± (0.3% from the measured value + 1.5 °C)
Resolution:
better than 0.1
°C (
to achieve this value, the
corresponding number of decimals must be set
)
Cold junction:
compensated in the temperature range of
(-
10 to 60) °C
Thermocouple of the “S” type (Pt10%Rh-Pt)
Measured quantity:
temperature
Range:
(0 to 1,700) °C
Accuracy without probes:
± (0.3% from the measured value + 1.5 °C)
Resolution:
better than 0.1
°C (
to achieve this value, the
corresponding number of decimals must be set
)
Cold junction:
compensated in the temperature range of
(-
10 to 60) °C
Thermocouple of the “B” type (Pt30%Rh-Pt)
Measured quantity:
temperature
Range:
(100 to 1,800) °C
Accuracy without probes:
± (0.3% from the measured value + 1 °C)
in the range of (300 to 1.800) °C
Resolution:
better than 0.1
°C (
to achieve this value, the
corresponding number of decimals must be set
)
Cold junction:
uncompensated