and cause false alarms.

 We recommend ventilating the room when products like these are used. Alcohols, Hydro-

gen and volatile organic solvents in great quantity cause false alarms, and then the cell needs one or more hours in 
clean air to recover the normal sensitivity conditions. 

TEST and CALIBRATION 

PAY  ATTENTION

This  procedure  has  to  be  made  with  extreme  attention  and  by  authorized  and  trained 

people;  because  starting  this  procedure  it  will  start  both  Outputs  (relays)  causing  the  activation  of  con-
nected alarm devices. 

Operation  Check,  Zero  Adjust,  and  Calibration

:

  are  different  code  protected  functions.  To  access  these  func-

tions  is  necessary  to  insert  the  relevant  “

Code

"  through  the  keys  F1s  and  F2.  To  have  the  key  pressure  recog-

nized, hold pressing it for around a second (

until the green Led doesn't switch off for a moment

). Then the next key can be 

pressed. In case of error all it takes is waiting around 10 seconds and the sequence is automatically erased. 
The 

Calibration Check

 is a function code free. 

Calibration Kit, Sample Gas Bottles

 (

for Calibration Check and Calibration

) the mixture to be used depending  on the 

model installed is shown in the table below: 

MODEL 

Sample gas bottle

 

Our Code / Valve 

KIT Code 

47RTS282EA - 47RTS282A-H 

100 ppm 

Ammonia (NH

3

in Nitrogen 

BO501

 / BO303

 

TC014 

47RTS282EC-S–47RTS282EC-H 

300 ppm 

Carbon monoxide (CO) 

in Nitrogen 

BO210

/ included

 

TC011

/TC014

 

47RTS282ES 

10 ppm 

Sulphur dioxide (SO

2

in Nitrogen 

BO518

/ BO303 

TC014

 

47RTS282EH 

50 ppm 

Hydrogen sulphide (H

2

S)  in Nitrogen 

BO520

 / BO303 

TC014

 

47RTS282EHCN 

10 ppm 

Hydrogen cyanide (HCN)  in Nitrogen 

BO529

 / BO303 

TC014

 

47RTS282EN 

100 ppm 

Nitric oxide (NO) 

in Nitrogen 

BO522

 / BO303 

TC014

 

47RTS282EN2 

30ppm 

Nitrogen dioxide (NO

2

in air 

Available on request 

TC014

 

47RTS282EHCL 

10ppm 

Hydrogen cloride (HCL) 

in Nitrogen 

BO527

 / BO305 

TC014

 

47RTS282ECL 

10ppm 

Nitrogen dioxide (NO

2

in air 

Available on request 

TC014

 

It is  possible  to  use either  the 

disposable cylinders  with adjust valve

 or the  high pressure  ones  with reduction 

gear.  Some  cylinders  have  a  fixed  flower  reducers,  make  the  necessary  value  in 

Table,  column  3,  page  11

.  Is  also 

necessary the 

calibration kit model TC014 (stainless steel)

. To ensure that the gas concentration reaches sen-

sor, use steel reduction gear and Teflon tubes or other compatible materials which do not disperse gas. 

“INSTRUMENT OPERATION CHECK” (

Check Code

: F2, F2, F1, F1)

this function allows to effect a functional test 

of the detector. After having inserted the "

Check C

ode

", all Led are switched off and all the relays are deactivated. 

Then they will switch on in sequence, the Led, from the yellow up to the 3° red. To the lighting of the various Led it 
corresponds the activation of the relevant relays (

relay "FAULT" with the yellow Led, relay "ALARM1" with the 1° red Led, relay 

"ALARM2"  with  the  2°  red  Led,  relay  "ALARM3"  with  the  3°  red  Led

).  At  the  end  all  the  Led  will  remain  lighted  for  around  5 

seconds,  then  the  detector  returns  at  the  conditions  of  normal  operation.  It  is  advisable  to  perform  this  operation 
every 12 months according to the use. 

Note

: this function is not working  if the  1° and/or the 2° and/or the 3° red 

Led have already turned on. 

“ZERO ADJUST” (

Zero Code

: F2, F1, F1, F2)

this function is to adjust the Zero sensor and can be done in clean 

air only (

environment without the presence of toxic gas or other pollutants

). Consider that the 

47RTS282E

 series have a so-

phisticated follower of zero, that every hour will reset the sensor, if of course there is the presence of gas. If neces-
sary,  perform  this  operation  after  installation  or  after  changing  the  “

Cartridge

”  or  every  6-12  months  depending  on 

environmental  conditions.  After  entering  the  "

Zero  Code

"  as  confirmation  of  the  operation,  there  will  be  a  flashing 

1

st

red Led and output becomes 4.0 mA. 

Note

: This function is not working if the 2

nd

 and/or 3th red Led are already ON or the mA output is higher than the 

value shown in the table below, depending on the model. In this case, please recalibrate the unit and/or replacing the 

Cartridge

”. In case the 1

st

 red Led will be already ON, only instead of flashing it will switch off for confirmation. 

47RTS282

EA 

47RTS282

EA-H

 

47RTS282EC-S

 

47RTS282

EC-H

 

47RTS282

ES 

47RTS282

EH 

47RTS282

EHCN 

47RTS282

EN 

47RTS282

EN2 

47RTS282

EHCL 

47RTS282

ECL 

4,8 mA 

15ppm NH

3

 

4,8 mA 

15ppm CO 

6,4 mA 

3ppm SO

2

 

4,8 mA 

5ppm H

2

S

 

8 mA 

2,5ppm HCN

 

5,6 mA 

10ppm NO

 

6,6 mA 

5ppm NO

2

 

8 mA 

2,5ppm HCL

 

4,8 mA 

0,5ppm HCL

 

“CALIBRATION” (

Calibration Code

: F2, F2, F2, F1, F2, F1)

this function allows completely recalibrate the sensor 

using the gas mixture indicated above.

 

Warning

to guarantee that no errors of elaboration happen, the possibility exists that during the Calibration the yel-

low Led switch off every 8 seconds, in this case interrupt the procedure, switch off and switch on the instrument and 
repeat the Calibration. If condition persists it will be necessary to send the detector to the supplier for the reparation. 

Important note: 

during Calibration routine the mA output indicates 0mA

.

 

The  “

Calibration

”  can  be  done  in  clean  air  only  (

environment  without  the  presence  of  toxic  or  other  polluting  gas

). With  the 

keys perform the “

Calibration Code

". Wait until the  yellow  and  green Led switch  on fix and the 1

st

 red Led starts to 

flash. Insert the TC014 on the sensor holder and adjust the influx of the gas the way that the flow meter reaches 
the value indicates in table below in 

column 3

, depending on the model (

see also Fig.3

). 

Wait around 3-5 minutes

, until the 3

rd

 red Led lights (

and while it is ON

), press the key F2 and hold it pressed until 

the 1

st

 and 3

rd

 red Led are switched off for at least 2 seconds (

but if the 1

st

 red Led continues to blink more than 5-6 

minutes means that the gas flow is not constant or unsatisfactory, or the concentration is not as required, stop the 
operation, turn off the instrument and recalibrate

). 

Then, close the gas cylinder and remove TC014

. At this point we 

can have two possibilities: 

Yellow  and  green  Led  illuminates: 

the  calibration  routine  is  correct. 

Wait  8  seconds,  until  the  instrument  auto-

matically restores the normal working conditions

(

See “

Operational Description > Preheating

). 

Summary of Contents for 47RT282EA

Page 1: ...rio Contents Sommaire Caratteristiche tecniche Technical specifications Caract ristiques techniques 2 IT DESCRIZIONE 4 NOTE SUL GAS RILEVATO DAI VARI MODELLI 4 FUNZIONAMENTO 5 INSTALLAZIONE 6 AVVERTEN...

Page 2: ...ensor limits Limite chelle 500 ppm 500 ppm 1000 ppm 1000 ppm 150 ppm 500 ppm Vita media in aria pulita Average Life in fresh air Vie moyenne en air pur 2 anni years ans 2 anni years ans 3 anni years a...

Page 3: ...th mois 2 mese month mois 2 mese month mois 3 mese month mois 2 mese month mois Risoluzione Resolution R solution 0 1 ppm 1 ppm 1 ppm 0 7 ppm 0 1 ppm Tempo di risposta T90 Response Time T90 Temps de r...

Page 4: ...respirato dalle persone dai polmoni raggiunge il sangue dove impedisce il trasporto dell ossigeno ai tessuti e agli organi 47RTS282EC H identico al precedente ma utilizza un sensore di livello superi...

Page 5: ...impostato Il rel ALARM 2 interviene dopo 30 secondi oppure immediatamente quando superato il livello TLV STEL Il 3 Led rosso ALARM 3 si accende se la concentrazione di gas supera il 3 livello d allarm...

Page 6: ...nati prima d alimentare il rilevatore Se lo si utilizza con i Dip Switch in posizione riservata vietata sar attivata l indicazione di Guasto vedi Funzionamento Guasti Importante terminata l installazi...

Page 7: ...ione del modello in questo caso procedere alla Calibrazione oppure sosti tuire la Cartuccia Se fosse gi acceso solo il 1 Led rosso invece di lampeggiare si spegner per conferma 47RTS282EA 47RTS282EA H...

Page 8: ...a toxic colourless and with a character istic pungent smell lighter than air its density as to air is 0 6 His TWA is 25ppm Time Weighted Average and STEL is 35ppm Short Term Exposure Limit His LEL Low...

Page 9: ...idity variations The cali bration is carried out with specific gas to be detected Anyway it can contemporaneously detect other gases that should be present in the same environment as listed in tables3...

Page 10: ...the conductors for output relay If the output signal is used in mA please use 3 conductors screened cables conductors for relay output The max load resistor is 50 ohm with 12Vdc power supply while is...

Page 11: ...ill remain lighted for around 5 seconds then the detector returns at the conditions of normal operation It is advisable to perform this operation every 12 months according to the use Note this functio...

Page 12: ...C sont d tecteurs ponctuel pour gaz toxiques quip e d un capteur a cellule lectrochimiques Le d tecteurs peut tre utilis en d tecteur ponctuel avec asservissements incorpor s ainsi qu en syst mes cen...

Page 13: ...rtie dans les proc d s de combustion en partie par les missions naturelles des ruptions volcaniques des incendies ou des activit s civiles et industriels de combustion tels que les transports v hicule...

Page 14: ...en panne la remplacer par une nouvelle 3 si une nouvelle Car touche est install e ou bien elle n est pas correctement connect e ou bien elle n est pas compatible Contr ler les connexions avec la Carto...

Page 15: ...de solvants organiques hydro g ne alcool en quantit s lev es des fausses alarmes peuvent appara tre et que le capteur devra pour retrouver ses conditions de fonctionnement normales rester longtemps en...

Page 16: ...et maintenir cette pres sion tant que la 1 re et la 3 me LED rouge ne restent pas teintes durant au moins 2 secondes si en revanche la 1 reLED rouge continue clignoter plus de 5 6 minutes cela signifi...

Page 17: ...0 10 2 Biossido di azoto Nitrogen dioxide dioxyde d azote 20 ppm 2 2 1 16 Cloro Clorine Chlore 1 ppm 0 0 0 Etanolo Ethanol thanol 200 ppm 0 0 2 5 0 Etilene Ethylene thyl ne 100 ppm 0 0 80 50 0 1 Idro...

Page 18: ...armes Relais Ritardo dei Rel in Secondi Relays Delay Seconds retarde du relais Secondes 1 2 3 ALARM 1 ALARM 2 ALARM 3 ALARM 1 ALARM 2 ALARM 3 FAULT OFF OFF OFF 1 5 ppm 2 5 ppm 5 0 ppm 12 30 60 30 ON O...

Page 19: ...vato Reserved R serv Risevato Reserved R serv OFF ON ON Risevato Reserved R serv Risevato Reserved R serv ON ON ON Risevato Reserved R serv Risevato Reserved R serv Tabella 12 Table 12 Tableau 12 47RT...

Page 20: ...a di collegamento Wiring diagram Sch ma TESTmV TP2 TP1 Scheda Uscite Rel Relays Output Board Sortie sur Realis S 12 24V mA 12 24Vdc Alimentazione Power Supply Alimentation Fig 3 Tester di calibrazione...

Reviews: