5.0 REAR
PANEL
INTERFACES
The rear panel has connectors for the data interface, fuse holder, security
lock location and power supply input.
5.1
POWER
Make certain your balance is compatible with the power supply in your area.
If the voltage is not the same the analyzer can be damaged. Contact your
supplier if you require assistance.
The power connection uses a standard IEC C13 socket on the cable with a
plug suitable for the location. Do not use adaptors if the plug does not match
your power supply.
The fuse should only be replaced with a time lag type fuse with high breaking
capacity, 5Amp current rating. The fuse is a standard 5 x 20mm size. To
replace the fuse remove all power from the analyzer, unscrew the fuse holder
cover to extract the old fuse, replace with a new fuse. If the fuse should blow
again, the analyzer must be serviced by a qualified service representative.
There are no user serviceable parts inside. Contact your supplier.
5.2 DATA
INTERFACE
The RS-232 interface is a general purpose serial interface for
communications with the analyzer, using a DE-9P type connector. Refer to
section 11.0 for details of the interface.
5.3 SECURITY
LOCK
The center of the rear panel has a location to mount a Kensington type
security lock. Refer to the instructions that come with the lock for details.