Page 20 of 60
6.1.2 PRINT CURRENT SETTINGS
When setting the test parameters the [Print/
] key will cause the current
parameters to be printed as shown below. However if the display is awaiting
a numeric entry the [Print/
] key will move the flashing digit to the left.
For example (English language version shown) :
CURRENT TEST SETTINGS
DATE: 08/01/2009
TIME: 14:44:49
SER NO:AE00001234
USER NO:1234567
Preset: PST 01
ABCD-01234
Result %MOISTURE
Heating: Single
157 C
Interval: 05 Sec
Stop: STABLE
0.002 g
15 Sec
Start: MANUAL
From the Real Time Clock
Analyzer Serial number
User ID number
If Preset test is used, blank if not
from a preset test.
Current test settings for heating
Printing interval
Stop Criteria
Start Criteria
6.2
STORE
TEST
SETTINGS
It will be possible to store up to 49 test procedures/settings. There is also a
factory default that is permanently stored as PST 00. See section 6.1
To Store the current settings press the [PST/
] key. The display will give
the user the option to store, recall or delete a Preset Test. The display will
show “PST STORE” or “PST RECALL” or “PST DELETE”. Use the up/down
arrow keys, [Dspl/ ] and [Test/ ], to select either store or recall.