4.5.6. Plant Status
On entering this option the plant status condition is displayed, i.e.
ON
or
OFF
.
Minutes, hours and days rolling averages are held and current outputs set to
zero, with alarms deactivated.
4.6. Normalisation
From this mode, the normalisation parameters that are currently being used can
be displayed. Press the
ENTER
key to enter the routine, and use the
ARROW
keys to select which of the normalising parameters to display.
Press the
MODE
key until the number 3 is seen in the top left corner of the
display. When the required normalising parameter is displayed, press the
ENTER
key to display the normalisation data. Press the
ENTER
key again to
exit the parameter.
The available parameters are :
•
temperature
•
oxygen
(wet or dry measured input)
•
pressure
•
water vapour