CODEL
OPS. 141
18
Issue : A
Revision : 1
Date : 14/06/18
Doc i/d :
0141/6
050028
9.
Ranges for the averages are :
minute 01 to 60 in 1 minute intervals
hour 01 to 24 in 1 hour intervals
day 01 to 30 in 1 day intervals
10.When days has been entered the LCD will read
RESET
, select
YES
and the display will prompt
SURE
. Select
YES
and after the
averages are reset the display will read
EXIT
. Press
ENTER
and
the processor will return to the
SET UP
mode menu option
AVERAGES
.
11.Use
DOWN
arrow to show
CONFIGURE O/P 1
(configure output
1). Press
ENTER
to access the menu. The first option is Output.
12.Press
ENTER
to access this option and use the
UP
and
DOWN
arrow keys to select either 0 or 4mA as the origin for current at the
outputs. Press
ENTER
when the correct one has been selected.
13.The
OUTPUT
option is exited and the
AVERAGES
option is
selected. Press
ENTER
to access the option and use
UP
and
DOWN
keys to select which average is to be displayed in the
operating mode.
14.Press
ENTER
and Units is the next option. Press
ENTER
to
access the option and use the UP and
DOWN
keys to select which
units are displayed. The choices are:
%OP
Optical Density
mg/m
3
mg/Nm
3
Ringelmann
15.Press
ENTER
and
Span
is displayed. Press
ENTER
and select
the span value required.
16.The first digit is selected by using the
UP
and
DOWN
arrow keys
and pressing
ENTER
, the cursor then moves to the second and
third digits in turn.
17.When all have been selected press
ENTER
.
18.The Span menu is exited and
Fault Cond
(condition) is viewed.
19.Press
ENTER
and select either :