CODEL
OPS. 141
45
Issue : A
Revision : 1
Date : 14/06/18
Doc i/d :
0118/6
050028
4.8.5.2. Units
This option is entered after Source has been completed. Select the units
required for the alarm as in section 4.8.3.3. Units. When these have been
selected press
ENTER
and the Level option is accessed.
4.8.5.3. Level
Select the required level for
ALARM 1
. Note that the current value is displayed
for 1 second, but then it defaults to zero and so must be re-entered for the
alarm level to be properly configured.
4.8.5.4. Exit
After the alarm level has been set press
ENTER
to reach this display. Press
ENTER
and the display exits to the
ALARM 1
option.
4.8.6. Alarm 2
Use the
DOWN
arrow key to select this after
ALARM
1 has been completed.
The parameters for the alarm for analogue output 2 are selected here in a
similar manner as for
ALARM 1
.
4.8.7. Parameters