22
Operation stops with
alarm:
Display shows
Power supply is
overheating
Ø
Turn off power supply. Check for
sufficient airflow around the power
supply fan. After cooling down, restart
the power supply by turning the Power
switch to the on position.
Ø
If you cannot restart the instrument,
turn off the power, disconnect the
power cord from the outlet, and
contact Technical Service.
Warning message
displays with 5-second
beep sound.
The screen shows
The power once been cut
and now recover
Ø
User does not need to take extra
action. The warning sign and beep
sound would only last for 5 second;
after that, the machine will continue
running the unfinished project.
Ø
The
sign indicates the machine
has been interrupted by sudden power
off. Press Enter Key
to clear the
sign.
Encountering Problems
1. Check the troubleshooting section.
2. Call Technical Service or e-mail to
3. If the unit must be shipped back for repair, contact Cleaver Scientific or the
distributor for a Return Material Authorization number and shipping instructions.
The unit will be repaired and returned to you as quickly as possible.