16
4.4
Program Mode
Instructions for operating PowerPro Series Power Supply in the programming
operation are provided in this section. The Program Mode allows you to vary levels in
voltage (V), current (mA), and power (W) during specified time periods for up to 6
Steps, depending upon your electrophoresis needs. The PowerPro Series power
supply is capable of having 30 different program files storages for user’s convenience.
After starting the operation (see the above 4.2 instruction), set the Program Mode as
follow:
Ø
Use
Key and
Key to select
Program
Mode
, and then press either
or
to
enter the next page.
Ø
Press
or
Key to select appropriate file
number, and then press
or
Key to
enter the next page.
Ø
Use
,
,
or
to move cursor
to parameters such as voltage (V) current (mA) or
power (W) or time (min.). Press
Key and
use
and
to set the appropriate value,
and then press
Key. Repeat this step until all
the parameters are set properly. There are 6
stages available for use; use
to move down
for setting Stage 4-6, and
to go back Stage
1-3.