Digital Power Meter 66203/66204 User’s Manual
E-6
The test result of power measurement:
66203/66204 Range
Voltage: 600V Range
Current: 20A Range
Fluke 5500 Output
Parameter
Max. Spec.
Measured
Result
Min. Spec.
480Vrms
16Arms
60Hz
PF = 1
W 7.6996k
7.6604k
PF
1
0.9981
66203/66204 Range
Voltage: 300V Range
Current: 5A Range
Fluke 5500 Output
Parameter
Max. Spec.
Measured
Result
Min. Spec.
240Vrms
4Arms
60Hz
PF = 1
W 962.46
957.54
PF
1
0.9981
66203/66204 Range
Voltage: 150V Range
Current: 2A Range
Fluke 5500 Output
Parameter
Max. Spec.
Measured
Result
Min. Spec.
120Vrms
1.6Arms
60Hz
PF = 1
W 192.49
191.51
PF
1
0.9981
66203/66204 Range
Voltage: 60V Range
Current: 0.5A Range
Fluke 5500 Output
Parameter
Max. Spec.
Measured
Result
Min. Spec.
48Vrms
0.4Arms
60Hz
PF = 1
W 19.249
19.151
PF
1
0.9981
66203/66204 Range
Voltage: 30V Range
Current: 0.2A Range
Fluke 5500 Output
Parameter
Max. Spec.
Measured
Result
Min. Spec.
24Vrms
160mArms
60Hz
PF = 1
W 3.8498
3.8302
PF
1
0.9981
66203/66204 Range
Voltage: 15V Range
Current: 0.05A Range
Fluke 5500 Output
Parameter
Max. Spec.
Measured
Result
Min. Spec.
12Vrms
40mArms
60Hz
PF = 1
W 481.23m
478.77m
PF
1
0.9981
Summary of Contents for 66203
Page 1: ...Digital Power Meter 66203 66204 User s Manual...
Page 2: ......
Page 3: ...Digital Power Meter 66203 66204 User s Manual Version 1 0 March 2014...
Page 8: ...vi...
Page 12: ......
Page 22: ......
Page 26: ......
Page 76: ......
Page 126: ......
Page 130: ......
Page 132: ......
Page 136: ......
Page 144: ......