Digital Power Meter 66203/66204 User’s Manual
5-28
Header off:
<NR2><RMT>
Example:
none
FETCh[:SCALar]:SIGMa:POWer:PFACtor?
MEASure[:SCALar]:SIGMa:POWer:PFACtor?
Type: Three-Phase.
Description:
These queries return the power factor. The return could be -3 or
<NR2>.
-3: Invalid data when OVR, OCR, OCP occur.
Setting syntax: none
Setting parameters: none
Query syntax:
FETCh:SIGMa:POWer:PFACtor?<PMT>,
MEASure:SIGMa:POWer:PFACtor?<PMT>,
Query parameters: none
Return parameters: <NR2>
Header on:
:FETCH:SIGMA:POWER:PFACTOR<space><NR2><RMT>
Header off:
<NR2><RMT>
Example:
none
FETCh[:SCALar]:SIGMa:POWer:APParent?
MEASure[:SCALar]:SIGMa:POWer:APParent?
Type: Three-Phase.
Description:
These queries return the apparent power. The return could be -3 or
<NR2>.
-3: Invalid data when OVR
、
OCR
、
OCP occur.
Setting syntax: none
Setting parameters: none
Query syntax:
FETCh:SIGMa:POWer:APParent?<PMT>,
MEASure:SIGMa:POWer:APParent?<PMT>,
Query parameters: none
Return parameters: <NR2>
Header on:
:FETCH:SIGMA:POWER:APPARENT<space><NR2><RMT>
Header off:
<NR2><RMT>
Example:
none
FETCh[:SCALar]:SIGMa:POWer:REACtive?
MEASure[:SCALar]:SIGMa:POWer:REACtive?
Type: Three-Phase.
Description:
These queries return the reactive power. The return could be -3 or
<NR2>.
-3: Invalid data when OVR
、
OCR
、
OCP occur.
Setting syntax: none
Setting parameters: none
Query syntax:
FETCh::SIGMa:POWer:REACtive?<PMT>,
MEASure:SIGMa:POWer:REACtive?<PMT>,
Query parameters: none
Return parameters: <NR2>
Header on:
:FETCH:SIGMA:POWER:REACTIVE<space><NR2><RMT>
Header off:
<NR2><RMT>
Example:
none
Summary of Contents for 66203
Page 1: ...Digital Power Meter 66203 66204 User s Manual...
Page 2: ......
Page 3: ...Digital Power Meter 66203 66204 User s Manual Version 1 0 March 2014...
Page 8: ...vi...
Page 12: ......
Page 22: ......
Page 26: ......
Page 76: ......
Page 126: ......
Page 130: ......
Page 132: ......
Page 136: ......
Page 144: ......