Programmable AC/DC Electronic Load 63800 Series Operation & Programming Manual
3-42
12. Press
2
,
0
then press
ENTER
to change the value to “00:20:00”.
I s e t = 1 8 . 0 0 0 C F = 1 . 5 0 0 P F = 0 . 8 0 0
I p ( m a x ) = 5 0 . 0 T I M E O U T = 0 0 : 2 0 : 0 0
C U T O F F V = 5 0 . 0 0 0
13. Move the cursor by pressing (
◄
,
►
) to the command line of “CUT OFF V=”.
I s e t = 1 8 . 0 0 0 C F = 1 . 5 0 0 P F = 0 . 8 0 0
I p ( m a x ) = 2 0 . 0 T I M E O U T = 0 0 : 2 0 : 0 0
C U T O F F V = 5 0 . 0 0 0
14. Press
1
,
1
,
0
then press
ENTER
to change the value to “110.00”.
I s e t = 1 8 . 0 0 0 C F = 1 . 5 0 0 P F = 0 . 8 0 0
I p ( m a x ) = 5 0 . 0 T I M E O U T = 0 0 : 2 0 : 0 0
C U T O F F V = 1 1 0 . 0 0 0
15. After setting the parameter of the load, the
LOAD ON/OFF
key should be pressed to turn
on the load, so that the load can sink the current. (For detailed information please refer
section 3.4.5
If there is error during the procedure of inputting value, press
CLEAR
to clear
the value at any time and then input again.
For switching to CF only mode, the procedures are as follows.
1. Press
CONF./LOCAL
key.
2. Press
1
, then press
ENTER
to enter SETUP.
3. Move the cursor (
▲
,
▼
) to “1-5. CF/PF SETTING”.
4. Press the direction key (
◄
,
►
) or rotate the rotary knob to choose “CF”, then press
ENTER
.
5. Then press
CC
to back to main page.
Users can press the arrow key to select the setting parameters on the LCD display, including
current RMS value, CF, I
p
(max), Timeout and Cutoff voltage. Users can type the values with
the Entry keys.