Remote Operation
4-15
Return Parameters: <CRD>, OFF, ON
[Unit = None]
Query Example:
DIO:OUT2?
CURRENT Subsystem
4.3.2.4
CURRent:STATic:L1
Type:
Channel-Specific
Description:
Set the static load current for constant current static mode.
Setting Syntax:
CURRent:STATic:L1<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:STAT:L1 20
Set the static load parameter L1 =
20A.
CURR:STAT:L1 10A
Set the static load parameter L1 =
10A.
CURR:STAT:L1 MAX
Set the static load parameter L1 =
maximum value.
CURR:STAT:L1 MIN
Set the static load parameter L1 =
minimum value.
Query Syntax:
CURRent:STATic:L1?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example:
CURR:STAT:L1?
CURR:STAT:L1? MAX
CURR:STAT:L1? MIN
CURRent:STATic:L2
Type:
Channel-Specific
Description:
Set the static load current for constant current static mode.
Setting Syntax:
CURRent:STATic:L2<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:STAT:L2 20
Set the static load parameter L2 =
20A.
CURR:STAT:L2 10A
Set the static load parameter L2 =
10A.
CURR:STAT:L2 MAX
Set the static load parameter L2 =
maximum value.
CURR:STAT:L2 MIN
Set the static load parameter L2 =
minimum value.
Query Syntax:
CURRent:STATic:L2?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example:
CURR:STAT:L2?
CURR:STAT:L2? MAX
CURR:STAT:L2? MIN
CURRent:STATic:RISE
Type:
Channel-Specific
Description:
Set the rising slew rate of current for constant current static mode.
Setting Syntax:
CURRent:STATic:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:STAT:RISE 2.5
Set rising slew rate to 2.5A/μs.
CURR:STAT:RISE 1A/μs
Set rising slew rate to 1A/μs.
CURR:STAT:RISE MAX
Set rising slew rate to the
maximum value of static load.
Summary of Contents for 63200A Series
Page 1: ...High Power Electronic Load 63200A Series Operation Programming Manual ...
Page 2: ...High Power Electronic Load 63200A Series Operation Programming Manual Version 1 0 March 2015 ...
Page 7: ...vi ...
Page 11: ......
Page 15: ......
Page 22: ...Overview 1 7 Figure 1 3 Dimension of 63200A with Protection Cover ...
Page 23: ......
Page 31: ......
Page 127: ......