High Power Electronic Load 63200A Series Operation & Programming Manual
4-20
RES:STAT:L2? MIN
RESistance:STATic:RISE
Type:
Channel-Specific
Description:
Set the rising slew rate of current for constant resistance mode.
Setting Syntax:
RESistance:STATic:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: RES:STAT:RISE 2.5
Set rising slew rate to 2.5A/μs.
RES:STAT:RISE 1A/μs
S
et rising slew rate to 1A/μs.
RES:STAT:RISE MAX
Set rising slew rate to the
maximum value of static load.
RES:STAT:RISE MIN
Set rising slew rate to the
minimum value of static load.
Query Syntax:
RESistance:STATic:RISE?[<space><MAX | MIN>]
Return Parameters:<NR2>,
[Unit = A/μs]
Query Example:
RES:STAT:RISE?
RES:STAT:RISE? MAX
RES:STAT:RISE? MIN
RESistance:STATic:FALL
Type:
Channel-Specific
Description:
Set the falling slew rate of current for constant resistance mode.
Setting Syntax:
RESistance:STATic:FALL<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: RES:STAT:FALL 2.5
Set falling slew rate to 2.5A/μs.
RES:STAT:FALL 1A/μs
Set falling slew rate to 1A/μs.
RES:STAT:FALL MAX
Set falling slew rate to the
maximum value of static load.
RES:STAT:FALL MIN
Set falling slew rate to the
minimum value of static load.
Query Syntax:
RESistance:STATic:FALL?[<space><MAX | MIN>]
Return Parameters:<NR2>,
[Unit = A/μs]
Query Example:
RES:STAT:FALL?
RES:STAT:FALL? MAX
RES:STAT:FALL? MIN
RESistance:STATic:IRNG
Type:
Channel-Specific
Description:
Set the current measurement range in CR mode.
Setting Syntax:
RESistance:STATic:IRNG<space><CRD | NR1>
Setting Parameters:<CRD | NR1>, LOW | L | 0, MIDDLE | M | 1, HIGH | H | 2
Setting Example: RES:STAT:IRNG HIGH
Set current range to High.
RES:STAT:IRNG M
Set current range to Middle.
RES:STAT:IRNG 0
Set current range to Low.
Query Syntax:
RESistance:STATic:IRNG?
Return Parameters:<CRD>, LOW(0), MIDDLE(1), HIGH(2)
[Unit = None]
Query Example:
RES:STAT:IRNG?
RESistance:DYNamic:L1
Type:
Channel-Specific
Description:
Set the load resistance during T1 period for constant resistance
dynamic mode.
Summary of Contents for 63200A Series
Page 1: ...High Power Electronic Load 63200A Series Operation Programming Manual ...
Page 2: ...High Power Electronic Load 63200A Series Operation Programming Manual Version 1 0 March 2015 ...
Page 7: ...vi ...
Page 11: ......
Page 15: ......
Page 22: ...Overview 1 7 Figure 1 3 Dimension of 63200A with Protection Cover ...
Page 23: ......
Page 31: ......
Page 127: ......