High Power Electronic Load 63200A Series Operation & Programming Manual
3-22
Time
Current
Load1
load2
10mS
10mS
10mS
10mS
4A
2A
T1
T2
T1
T2
Figure 3-6 Dynamic Current Waveform
CRD Mode
3.6.6
In CRD mode, the Load will sink a dynamic resistance according to the programmed
resistance and dynamic timing by the input voltage. To enter into the CRD mode, press the
MODE
key and select
CRD
mode.
L 1 =
1 0 0 0 . 0 0
Ω
T 1 =
1 0 0 . 0 0
m s
C C
L 2 =
1 0 0 0 . 0 0
Ω
T 2 =
1 0 0 . 0 0
m s
▼
C R
0 . 0 0 0 V
0 . 0 0 0 A
C V
0 . 0 0 0
V p +
0 . 0 0 0
V p -
0 . 0 0 0 0
W
C P
C R D _ H
V o n
=
0 . 0 0 0
V
S R
=
0 . 0 0 5
A / u s
R E P E A T =
0
C C
S R
=
0 . 0 0 5
A / u s
I _ R A N G E =
H I H G ( 2 )
▼
C R
0 . 0 0 0 V
0 . 0 0 0 A
C V
0 . 0 0 0
V p +
0 . 0 0 0
V p -
0 . 0 0 0 0
W
C P
C R D _ H
V o n
=
0 . 0 0 0
V
Parameters:
L1: Set the loading value for Load1.
L2: Set the loading value for Load2.
SR
↗
: Set the current rise slew rate data.
SR
↘
: Set the current fall slew rate data.
T1: Set the loading time for L1.
T2: Set the loading time for L2.
Summary of Contents for 63200A Series
Page 1: ...High Power Electronic Load 63200A Series Operation Programming Manual ...
Page 2: ...High Power Electronic Load 63200A Series Operation Programming Manual Version 1 0 March 2015 ...
Page 7: ...vi ...
Page 11: ......
Page 15: ......
Page 22: ...Overview 1 7 Figure 1 3 Dimension of 63200A with Protection Cover ...
Page 23: ......
Page 31: ......
Page 127: ......