Operation Overview
3-27
CZ Mode
3.7.5
In CZ mode, the Load will sink a current according to the programmed impedance.
Impedance can be programmed by set the equivalent series resistance Rs, equivalent series
inductance Ls, equivalent parallel load capacitance C
L
, equivalent parallel load resistance R
L
and Ip (max)
parameters for loading when operating in this mode.
To enter into CZ mode, press
and select
CZ
and then press Enter.
Figure 3-2 Constant Impedance Mode
Parameters:
C
L
: Set the level of equivalent parallel load capacitance C
L
. The range is 30
μF to 50,000μF.
R
L
: Set the level of equivalent parallel load resistance R
L
. The range is the same as the CR
mode high range.
Ls: Set the level of equivalent series inductance Ls. The range is 0.1
μH to 20μH.
Rs: Set the level of equivalent series resistance Rs. The range is
30mΩ to 20Ω.
CAUTION
It is suggested to use a remote sense cable to measure the UUT output
voltage.
CVCC
3.7.6
In CVCC mode, the Load will adjust the sink current to control the output voltage of current
source by the programmed voltage. Constant voltage has three types of response speed:
fast, normal and slow.
To enter into CVCC mode, press
and select
CVCC
and then press Enter.
V
I
CV
CC
CV+CC mode:I/V Curve
Parameters:
V_SET: Set the voltage level.
ADVA
ADVA
Summary of Contents for 63200A Series
Page 1: ...High Power Electronic Load 63200A Series Operation Programming Manual ...
Page 2: ...High Power Electronic Load 63200A Series Operation Programming Manual Version 1 0 March 2015 ...
Page 7: ...vi ...
Page 11: ......
Page 15: ......
Page 22: ...Overview 1 7 Figure 1 3 Dimension of 63200A with Protection Cover ...
Page 23: ......
Page 31: ......
Page 127: ......