Programmable Bidirectional DC Power Supply 62000D Series
Operating and Programming Manual
3-20
When Foldback protection occurs the main page will prompt a protect message as
Figure 3-29 shows.
Figure 3-29
Be aware that if delay time sets to
t
seconds, it means the Foldback that set to CV TO
CC or CC TO CV won’t be activated unless it sustains
t
seconds when a mode change
is detected. If the change time of mode is less than
t
seconds it will return to its original
state and Foldback protection will not occur as Figure 3-30 shows.
DELAY TIME
C V MODE
CC
MODE
Time
FOLDBACK
PROTECTION
DC SOURCE OPERATION MODE
Figure 3-30
Assuming the Foldback is set to CV TO CC, the solid line in Figure 3-30 will create
Foldback protection while the dot line will not.
3.2.3.5
Meas. Setting
1.
Tap Meas. Setting to enter into Measurement Setup as Figure 3-31 shows.
2.
There are Average Time and Average Method for setting.