Manual Operation
3-5
Figure 3-5
3.2.1.3
Load Mode
The following describes how to set Load Mode:
1.
Tap the drop-down menu and select Load Mode; the panel will switch to the screen as
Figure 3-6 shows.
2.
The rest of the settings are same as Source & Load Mode.
In Load Mode, positive current will still be generated even when the
output voltage is drops suddenly. (It is recommended to add a blocking
diode at customer site.)
WARNING
Load Mode
DUT
62000D
Bi-DC SOURCE
SVO+
VO+
SVO-
VO-