Remote Operation
5-33
19. SYSTem:COMMunicate:SOCK:IP
Description:
Sets the Ethernet IP.
Syntax:
SYSTem:COMMunicate:SOCK:IP <Arg>
Parameter:
<Arg>: “192.168.1.1”
Example:
SYSTem:COMMunicate:SOCK:IP “192.168.1.1”
Query Syntax:
SYSTem:COMMunicate:SOCK:IP?
Return Parameter:
“192.168.1.1”
Query Example:
SYSTem:COMMunicate:SOCK:IP?
Return Example:
“192.168.1.1”
20. SYSTem:COMMunicate:SOCK:MASK
Description:
Sets the Ethernet IP Mask.
Syntax:
SYSTem:COMMunicate:SOCK:MASK <Arg>
Parameter:
<Arg>: “192.168.1.1”
Example:
SYSTem:COMMunicate:SOCK:MASK “192.168.1.1”
Query Syntax:
SYSTem:COMMunicate:SOCK:MASK?
Return Parameter:
“192.168.1.1”
Query Example:
SYSTem:COMMunicate:SOCK:MASK?
Return Example:
“192.168.1.1”
21. SYSTem:COMMunicate:SOCK:APPLY
Description:
Updates the Ethernet setting.
Syntax:
SYSTem:COMMunicate:SOCK:APPLY
Parameter:
None
Example:
SYSTem:COMMunicate:SOCK:APPLY
Query Syntax:
SYSTem:COMMunicate:SOCK:APPLY?
Return Parameter:
DONE | UNDONE
Query Example:
SYSTem:COMMunicate:SOCK:APPLY?
Return Example:
DONE | UNDONE
5.6.2.9
INSTRUMENT Subsystem
1.
INSTrument:STATus:AD?
Description:
Returns the AD module status.
Syntax:
INSTrument:STATus:AD? <NR1>
Query Syntax:
1 | 2 | 3
Return Parameter:
0 ~ 4294967295 (2
32
-1)
BIT 0
AD_VDC_OVP
BIT 16
DD_IO_REG_OCP
BIT 1
AD_VDC_UVP
BIT 17
AD_RLY_STARTFAIL
BIT 2
AD_VRS_OVP
BIT 18
AD_PWM_TOP_FAULT
BIT 3
AD_VTR_OVP
BIT 19
AD_PWM_BOT_FAULT
BIT 4
AD_VST_OVP
BIT 20
AD_AC_STARTFAIL
BIT 5
AD_VRS_UVP
BIT 21
AD_PFC_STARTFAIL
BIT 6
AD_VTR_UVP
BIT 22
AD_HARD_ERR
BIT 7
AD_VST_UVP
BIT 23
DD_VO_UVP_S
BIT 8
DD_VO_OVP_F
BIT 24
AD_MODEL_RES_ERR
BIT 9
DD_VO_UVP_F
BIT 25
DD_SHORT
BIT 10
AD_IR_OCP
BIT 26
AD_MEM_ERR
BIT 11
AD_IT_OCP
BIT 27
DD_LLC_STARTFAIL
BIT 12
AD_IS_OCP
BIT 28
AD_VAC_UBL
BIT 13
AD_Vd_OVP
BIT 29
DD_IP_OCP
Summary of Contents for 62000D Series
Page 1: ......
Page 2: ...Get more information by downloading Chroma ATE Solutions APP ...
Page 8: ...vi ...
Page 9: ...vii ...
Page 10: ...viii ...
Page 14: ......
Page 18: ......
Page 35: ...Overview 1 17 Figure 1 6 Dimension of 62000D 600 1200 1800 Models ...
Page 36: ......
Page 52: ......
Page 63: ...Manual Operation 3 11 Figure 3 13 Figure 3 14 ...
Page 109: ...Manual Operation 3 57 _EXT ON SIGNAL EXT ON DISABLE EXT ON ENABLE Output Output Figure 3 85 ...
Page 110: ......