5.2
Supply and disposal connections
Pos : 140 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/050 Aufst ellung und Ansc hluss/ 050 Aufst ellung und Ans chl uss Epsilon/ 050-0020- 0010 Energiev ers orgung Epsilon --------------------------------------- @ 48\mod_1490087247571_68. docx @ 492266 @ 3 @ 1
→
5.2.1
Power supply
Pos : 141 /200 Chris t/361 GT-BA Labor-Pil ot ( PROJ EKTE)/ Epsilon 6er/ St andard/ 050 Aufs tell ung und Ansc hluss /050-0020-0010- 0010 Ansc hlus sart E2-6D @ 25\mod_1404984829629_68. docx @ 186267 @ 4 @ 1
→
5.2.1.1
Connection
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage
WARNING
Work on the power supply system must only be performed by certified
electricians.
Christ freeze-dryers are units of safety class I. Epsilon 2-6D LSCplus units
have a five-wire power cord with a 5-pin CEE-plug (see chapter 10 -
"Technical data").
CAUTION
Prior to connecting the unit to the power supply system, check the rotary
field at the electrical outlet with a suitable measuring instrument:
The rotary field must be clockwise!
Pos : 142 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 143 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/050 Aufst ellung und Ansc hluss/ 050 Aufst ellung und Ans chl uss Epsilon/ 050-0020- 0010-0020 Sic herungen baus eits Epsilon @ 25\m od_1404984623550_68.docx @ 185665 @ 4 @ 1
→
5.2.1.2
Customer-provided fuses
The freeze-dryer must be protected by way of a customer-provided mains
fuse with the correct rating.