Freeze-dryer Epsilon 2-6D LSCplus
7 Malfunctions and error correction
Version 07/2022, Rev. 3.0 of 20/07/2022
• sb
113 / 156
Translation of the original operating manual
Pos : 225 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020-0022 Bel üft ungs-, Medienabl aufv entil LSCpl us Epsilon (auß er 1-4/ 2-4) @ 36\mod_1436511271936_68. docx @ 267302 @ 4 @ 1
→
7.1.2.2
Aeration valve, media drain valve
WARNING
Risk of poisoning/infection caused by contaminated condensate
(defrosting water)
The condensate may contain harmful substances originating from the
product.
Contact with the condensate may cause severe damage to health.
• Wear suitable protective clothes, gloves, and respiratory protection
when maintaining the drain system (especially when cleaning the
valves and when replacing the seals)!
A malfunction of the aeration valve or the media drain valve may have
several causes. One potential source are contaminants such as product
residues within the valve.
• Switch the freeze-dryer off and disconnect the mains plug.
• Clean the valve (see chapter 8.1.4 - "Aeration valve, media drain
• Put the freeze-dryer into operation again.
If there is still a leakage, the freeze-dryer must be checked by qualified
specialist personnel (see chapter 7.3 - "Service contact").
Pos : 226 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 227 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020-0025 Druckst euerv entil @ 25\m od_1404984063525_68. docx @ 184209 @ 4 @ 1
→
7.1.2.3
Pressure control valve
A malfunction of the pressure control valve may have several causes.
NOTE
The inspection of the valve must be carried out by qualified specialist
personnel (see chapter 7.3 - "Service contact").
Pos : 228 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 229 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020-0040 Vak uummess sonde allgem ein @ 25\mod_1404984065691_68. docx @ 184237 @ 4 @ 1
→
7.1.2.4
Vacuum sensor
Vacuum sensors have a limited service life and can be ordered as spare
parts.
Capacitive vacuum sensors
Capacitive vacuum sensors may experience a measurement drift due to
long-term use, soiling or sudden aeration. In this case, the vacuum sensor
must be adjusted (see chapter 8.1.11 - " Vacuum sensor ").
Pos : 230 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1