Freeze-dryer Epsilon 2-6D LSCplus
7 Malfunctions and error correction
112 / 156
Version 07/2022, Rev. 3.0 of 20/07/2022
• sb
Translation of the original operating manual
Pos : 221 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020 Kein ausr eichendes Vakuum @ 25\m od_1404984060039_68.docx @ 184167 @ 3 @ 1
→
7.1.2
Insufficient vacuum
NOTE
The vacuum checks must be carried out when the ice condenser is
frozen.
Pos : 222 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 223 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/070 St örungen und F ehl ersuche/ 070- 0010- 0020-0010 Klei nflansc hv erbi ndungen @ 25\m od_1404984061086_68. docx @ 184181 @ 4 @ 1
→
7.1.2.1
Small flange connections
Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
• Loosen the connection and place the centring ring (with sealing ring
inside) in a centred manner between the flange connections.
• Seal the connection with the clamping ring by tightening the wing nut.
• Ensure that the centring ring neither slips out of place nor gets jammed.
Fig. 69: Small flange and centring ring
Fig. 70: Small flange with centring ring
and small flange
Fig. 71: Attaching the clamping ring
Fig. 72: Tightened clamping ring
Pos : 224 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1