Freeze-dryer Epsilon 2-6D LSCplus
5 Set-up and connection
52 / 156
Version 07/2022, Rev. 3.0 of 20/07/2022
• sb
Translation of the original operating manual
Pos : 155 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/050 Aufst ellung und Ansc hluss/ 050 Aufst ellung und Ans chl uss Epsilon/ 050-0020- 0060 Drucks teuerventil St andard 2. 0 ( auß er 2-4) @ 48\mod_1490093015317_68. doc x @ 492528 @ 3 @ 1
→
5.2.5
Pressure control valve
The pressure control valve is integrated in the suction pipe between the
vacuum pump and ice condenser chamber. During certain, specified
process phases, it interrupts the volume flow to the vacuum pump (see
chapter 2.2.1 - "General information on freeze-drying")
CAUTION
The pressure control valve must be installed in the direction of the volume
flow in the indicated installation direction (see illustration below)!
1
Installation direction
of the pressure
control valve
Fig. 11: Vacuum pump with pressure control valve (example illustration)
Pos : 156 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1