Freeze-dryer Epsilon 2-10D LSCplus
6 Operation
Version 07/2022, Rev. 3.0 of 20/07/2022
• sb
83 / 161
Translation of the original operating manual
If the freeze-dryer is to be extended by an option, this option must be
enabled via this dialog box.
• Press the button "Add option". An input window opens.
• Enter the six-digit Christ activation code that was supplied for this
option. Note that the keys are case sensitive.
Options can be removed in the same way.
NOTE
The modifications will not become effective until after a restart of the unit.
Configuration of sensors (6)
CAUTION
The sensor configurations depend on the equipment version of the unit. If
the configuration is incorrect, the correct operation of the unit cannot be
guaranteed.
This dialog box is used for the configuration of the existing sensors.
Fig. 44: Dialog box "Configuration of sensors" (example
Vacuum sensors (see the illustration above, item A)
For the vacuum as well as pressure, there are two different measuring
methods, and for each of these methods, different sensors can be selected.
• In order to change the sensor, press the button with the sensor name.
The possible models will be displayed one after the other.
The buttons on the right-hand side of the sensors show a control symbol.
The tick marks on the buttons indicate the control sensors, i.e. the sensors
that are decisive for the vacuum inside the unit. If there is a second sensor,
the data of this sensor are simply used for comparison.