42/44
EN
Kinetec Kinevia™ / Kinetec Kinevia Duo™
Information about electromagnetic compatibility
All the information given below comes from standard requirements to which electromedical devices are subject as
defined by the IEC60601-1-2 standard.
The device radio-frequency emissions are very low and are not therefore likely to cause interference with electronic
equipment installed nearby (radios, computers, phones, etc.). Nevertheless, users will make sure that possible
electromagnetic interferences do not create an additional risk, such as from radio-frequency emitters or other
electronic devices.
The device is designed to withstand foreseeable interferences from electrostatic discharges, magnetic fields from
mains power supplies or radio-frequency emitters. Nevertheless, some types of mobile telecommunications devices
such as mobile phones may interfere with the medical device. The separation distances recommended in this section
must therefore imperatively be respected.
In this section, you will find the information required to guarantee the installation and commissioning of your medical
device in the best conditions in terms of electromagnetic compatibility.
WARNING
MALFUNCTION RISK
If there is mutual electromagnetic or other interference with other devices,
move the device away from them.
The device must not be used next to other devices or stacked with them
as this could lead to malfunctions. If this use is required, observe this and
the other devices to check that they are still in working order.
●
Cable list
List of cables likely to alter device compliance with electromagnetic emissions and immunity:
Description
Kinetec reference Maximum length
Europe mains cord
4610009092
3.50m
US/Japan mains cord
4610009109
4.00m
Brazil mains cord
4610009117
3.50m
UK mains cord
4610009125
3.50m
Australia mains cord
4610009133
3.50m
Swiss mains cord
4610009365
3.50m
WARNING
MALFUNCTION RISK
The use of accessories and cables other than those indicated or supplied
by KINETEC SAS can cause increased electromagnetic emissions or
reduce device immunity and cause malfunctions.
●
Recommended separation distances
The device is designed for use in an electromagnetic environment in which radiated radiofrequency interference
is controlled.
WARNING
MALFUNCTION RISK
Portable radio-frequency communications devices (including peripherals
such as aerial cables and external aerials) should not be used closer than
30 cm (12 inches) from any part of the device, including the cables
specified by Kinetec. If this is not the case, the performances of those
devices could be altered.