Training duration
You can set the durations for the different training phases for each training level using this menu. The following values
are set by default.
Level
Warm up
(min 2 / max 60)
Training
(min 2 / max 60)
Recovery
(min 2 / max 60)
2 minutes
6 minutes
2 minutes
2 minutes
10 minutes
3 minutes
2 minutes
19 minutes
4 minutes
Training settings
You can set the default training parameters for each program, each training level and each training phase using this
menu.
Date and time
You can set the date and time on this menu using the arrows to move from field to field and the plus/minus keys to
change the values.
The date and time are saved when the Kinevia switches to standby mode.
The date and time must be reset if the Kinevia has been powered off.
User data
This menu is used to set patient data using the arrows to move from field to field and the plus/minus keys to change
values such as age, height, weight and gender.
This data is required to calculate the exerted power (Watts) and energy burned (calories).
Technical settings
Different Kinevia settings can be adjusted using this menu.
◦
Rpm limit:
is used to define a pedalling rotation speed not to be exceeded (min 10 / max 60).
◦
Lock changes:
when this option is enabled, all the keys are locked during training except the “stop” key. The
symbol is displayed at the top right of the screen.
◦
Allow changes:
when this option is enabled, you can choose to save the settings used as default settings for
the next training sessions.
◦
Factory settings:
is used to reset all the Kinevia settings to their default values, such as duration and training
settings (see sections 0 and 12.2), statistics (see section 0
), …