![Chauvin Arnoux C.A 6160 User Manual Download Page 37](http://html1.mh-extra.com/html/chauvin-arnoux/c-a-6160/c-a-6160_user-manual_2594583037.webp)
37
How to operate REMOTE CONTROL pedal
START/STOP function on the pedal is exactly the same as on front panel of the instrument when the pedal is not connected.
SAVE function on the pedal is automatic, so a double press to SAVE pedal is required to save displayed result to the next location
of set device number. The device number must be set in advance. If you press SAVE more than twice, the instrument will throw into
Recall function and you cannot exit it by using REMOTE PEDAL. You can exit this function only by pressing front panel key on the
instrument.
The following procedure is to be used:
STEP 1. Connect REMOTE CONTROL pedal to the instrument as it is shown in fig.56 and carry out the measurement by pressing
START/STOP pedal.
STEP 2. Save the first result to desired memory location (memory number and device number) using front panel keys, see the
instructions in chapter 5.2.
STEP 3. Carry out the next test using START/STOP pedal.
STEP 4. Save the result by pressing SAVE pedal twice.
STEP 5. Proceed with the measurements.
Technical specifications of the pedal:
■
Cable length
10 m
■
Commands
START/STOP, SAVE
■
Casing
metal
■
Weight
2 kg
■
Size (W
×
H
×
D)
(300
×
55
×
175) mm
5.8. USE OF WARNING LAMP
The lamp is intended to inform the user if dangerous voltages are present when WITHSTANDING tests are carried out (HV and
PROG.HV position).
Meaning of each lamp:
-
Red (TEST) lamp on means that dangerous voltage is present at WITHSTANDING test terminals. Take care when
using test pistols!
-
Green (READY) lamp on means that the instrument is ready for the next measurement, dangerous voltage is not
present at WITHSTANDING test terminals.
1
2
3 4
5
T ES T
F 6
RE A DY
12V DC
-
+
Instrum en t con nector
(fem a le) typ e M A B 5 100
D IN 4 152 4 (front view )
an d in terna l instrum e nt
circu itry
fro nt view
Fig. 57. Connection of warning lamp to APPLIANCE MULTITESTER
Technical specifications of the WARNING LAMP:
■
Cable length
1 m
■
Lamp bulbs
(12 - 15) V / 4 W
■
Casing
plastic
■
Weight
0.3 kg
■
Size (W
×
H
×
D)
(200
×
95
×
110) mm
NOTE !
If no lamp lights when rotary switch is in HV position, stop the measurements immediately and check the WARNING LAMP
connection and light bulbs.