Frequency readout
Range (Hz)
Resolution (Hz)
Accuracy
45.00 – 65.00
0.01
±
(0.1% of r3 dig.)
Threshold Apparent Power: ................................. adjustable
(10 - 3500) VA
(10 - 100) VA (resolution 1 VA)
(100 - 3500) VA (resolution 10 VA)
Output: ................................................................. 16 A power test socket
Timer: ................................................................... adjustable 1 s - 9 min 59 s with resolution 1 s
Timer OFF function available.
3.8. DISCHARGE TIME (DISC.TIME POSITION)
■
Discharge time on power plug (external)
Max. working voltage ....................................................... 600 V p
Min. working voltage ........................................................ (60, 120) V p
Measuring range ............................................................. (0 - 10) s
Resolution ....................................................................... 0.1 s
Threshold of discharging time ......................................... 1 s
Accuracy .......................................................................... ±(2 % of r 0.2 s)
Safe voltage level ............................................................ 60 V, 120 V
Internal resistance of input .............................................. 96 M
Ω
■
Discharge time on internal electronic components (internal)
Max. working voltage ....................................................... 600 V p
Min. working voltage ........................................................ (60,120) V p
Measuring range ............................................................. (0 - 10) s
Resolution ....................................................................... 0.1 s
Threshold of discharging time ......................................... 5 s
Accuracy .......................................................................... ±(2% of r 0.2 s)
Safe voltage level ............................................................ 60 V, 120 V
Internal resistance of input .............................................. 96 M
Ω
3.9. GENERAL
Mains voltage: ................................................................. 230 V (-10 % - +6 %) / (50,60) Hz
Max. power consumption: ................................................ 660 VA (without load on TEST SOCKET)
Display: ............................................................................ LCD dot matrix, (160 x 116) dots with cont. backlight
RS232 interface: .............................................................. 1 start bit, 8 data bits, 1 stop bit
RS232 baud rate adjustable: .......................................... 9600, 19200, 38400 Baud
Memories: ........................................................................ 1638 memory locations
Remote control signals: ................................................... START / STOP, SAVE
EXT/ Door in signals: ....................................................... Next test, Good / Bad result, External input, Door in
Bar code reader: .............................................................. EAN13
Bar code baud rate adjustable: ....................................... 2400, 4800 or 9600 Baud
Measurement circuitry protection:
F3
.................................................................................. F 2.5 A / 250 V (5
×
20) mm (general protection of the instrument)
F4
.................................................................................. F 2.5 A / 250 V (5
×
20) mm (general protection of the instrument)
NOTE !
For correct operation of the instrument fuses F3 and F4 must be in good condition because the test socket is the point from
which the instrument observes the input voltage (zero crossing for soft start of internal generator).
Case: ............................................................................... shock proof plastic / portable
Dimensions (w
×
h
×
d): ................................................... (410
×
175
×
370) mm
Mass (without accessories): ............................................ 13.5 kg
Pollution degree: ............................................................. 2
Degree of protection (at closed cover): ........................... IP 50
Overvoltage category: ..................................................... Cat III / 300V, Cat II / 600V
Protection classification: .................................................. I
Working temp. range: ....................................................... (0 - +40)
0
C
Ref. temp. range: .............................................................. (+5 - +35)
0
C
Ref. humidity range: ......................................................... (+40 - +70) % RH
Storage temp. range: ....................................................... (-10 - +60)
0
C
Max. working humidity: .................................................... 85% RH (0 - +40)
0
C
Max. storage humidity: ..................................................... 90% RH (-10 - +40)
0
C
80% RH (+40 - +60)
0
C
10