15
P ro g k e y to s e le ct p ro g ra m in g m e n u
T rip p in g o u t c u rre nt
Fig. 10. Main menu in PROG.HV function
STEP 2. Press Prog. key to set or to check programmed ramp values in order to prevent damage to the equipment under test (last
values used are memorized). Select test parameters as follows:
Fig. 11. Menu for programming ramp values
■
Press Ilim key in order to reach menu for trip-out current selection and character of displayed part of leakage current
(resistive or capacitive). The same procedure as for Ilim selection at HV function applies.
To change U and T values press U/T key. Value T2 represent time from t1 to t2 and value T3 represent time from t2 to t3. (fig. 12).
Menu for selection and changing values is displayed, see the following figure:
Fig. 12. Timer T1 is selected, change value by using
↑ ↓
↑ ↓
↑ ↓
↑ ↓
↑ ↓
keys
-
To select ramp times or voltage values press Sel. key.
-
Use
↑
and
↓
keys to set appropriate test value:
-
Time: (1 s - 240 s)
-
Voltage: 100 V –5 kV
-
Press Exit key (2x) to exit.
STEP 3. Connect test probes (pistols) to the instrument.
STEP 4. Close DOOR IN safety connector, if enabled. (CONTINUITY test terminals must be open).
STEP 5. Press START/STOP key to start high voltage generator and carry out the test using test probes.
STEP 6. Wait for the test time to elapse or press START/STOP key again to stop high voltage generator.
STEP 7. Save displayed result for documentation purpose (see instruction in chapter 5.2. on how to save displayed result).