Carestream CS 8100 3D User Manual Download Page 87

Panoramic and 3D Modality User guide for CS 8100 3D Family (SM842) Ed07

81

 

appears

 

in the Acquisition Export Mode field.

7.

Select either one of the appliance acquisition options.

Launching the X-ray

To launch the X-ray, follow these steps:

1.

Leave the X-ray room and close the door. 

2.

On the 

Acquisition

 interface, when 

 is green, press and hold the exposure button of the X-ray 

remote control until the end of acquisition. 

The 

 indicator in the acquisition interface turns yellow and a warning sound is emitted to 

indicate that X-ray emission is in progress.
The acquired image appears in the 

Preview Screen

3.

Once all items have been scanned, the 

Acquisition

 interface closes. Wait for the 3D object 

reconstruction.

The reconstruction object will display in the image browser.

Important:  The animated display screen will guide you on how to 
position the acquisition mode that you have selected.

WARNING:  Make sure that you position the acquisition material 
correctly to obtain quality data.

Important:  If you have a problem that requires you to stop the 
acquisition, release the exposure button of the remote control or 
press the red emergency stop button.

Summary of Contents for CS 8100 3D

Page 1: ...User Guide CS 8100 3D Family CS 8100 3D CS 8100 3D Access CS 8100SC 3D CS 8100SC 3D Access...

Page 2: ...ity without the 26x24 Field of View FoV The system has the capability to expand the field of view to the 3D full upper and lower jaw acquisition and to the cephalometric 26x24 FoV with licence upgrade...

Page 3: ...ne 15 3D Acquisition Interface Overview 16 3D Program Pane 17 3D Patient Pane 20 3D Parameter Pane 21 Chapter 4 Getting Started Switching on the Unit 23 Increasing the operating life of the X ray tube...

Page 4: ...etting the Acquisition Parameters 62 Launching the Acquisition for the Impression Material 63 Launching the Triple Tray Impression Acquisition 65 Acquiring Images of CS Restore Plaster Objects 67 Prep...

Page 5: ...To Acquire an Image of the NobelGuide or NobelClinician Calibration Phantom 84 Chapter 8 Maintenance Maintenance actions 85 Monthly maintenance 85 Annual maintenance 85 Controlling the Image Quality 8...

Page 6: ...vi...

Page 7: ...rmation or indicate potential risk to personnel or equipment WARNING Warns you to avoid injury to yourself or others by following the safety instructions precisely CAUTION Alerts you to a condition th...

Page 8: ...2 Chapter 1 Conventions in This Guide...

Page 9: ...n and 3D full upper and lower jaw acquisition and cephalometric modality see relevant documentation CS 8100SC 3D Access Panoramic modality dental volumetric reconstruction modality limited to 3D focus...

Page 10: ...ents Figure 1 illustrates the Up and down movement of the unit Rotation and translation movement of the rotative arm Figure 1 CS 8100 3D Unit Mobile Components Important The patient can enter by eithe...

Page 11: ...al functional components of the unit Figure 2 Unit Functional Components 1 ON OFF button 7 Temple support 2 Emergency stop knob 8 Digital sensor 3 Unit rotative arm 9 Unit head 4 X ray source assembly...

Page 12: ...llowing figure illustrates the functional components of the head and chin rest Figure 3 Head and Chin Rest Functional Components 1 Positioning Panel 4 Bite block support 2 Temple support adjustment kn...

Page 13: ...ent using the unit Table 1 Positioning Accessories Accessory Description Panoramic chin rest Sinus chin rest TMJ nose rest Standard bite block x5 Frankfort guide bite block for panoramic x3 Bite block...

Page 14: ...bite block x4 2 3D molar bite block x2 4 3D child bite block x2 2 3rd Molar offset 3D bite block x2 for use with 5x5 Ml and Mr FoV only Temple support cone for use with 3rd Molar offset 3D bite block...

Page 15: ...e height of the patient When the unit is not in use places the rotative arm in a parallel position to the unit head leaving more free space around the unit To do so press and hold both buttons until t...

Page 16: ...e the X ray room You must press and hold the exposure button until the end of acquisition Premature release of the exposure button interrupts the acquisition Figure 5 X ray Remote Control 1 Exposure b...

Page 17: ...sentations of guided surgical images for double scan protocol producing acquisition files that are compatible with competitor solutions For user instructions see Chapter 7 3D Object Acquisition Imagin...

Page 18: ...r warning messages originating from the unit 5 Generator Cooling indicator Indicates the automatic cooling time mm ss required before the next acquisition 6 Reset button Resets the unit to the patient...

Page 19: ...diological exam options Full panoramic acquisition Click on all the areas of interest Example Segmented panoramic acquisition Click on the following areas of interest that you want to examine Posterio...

Page 20: ...nd bone density of the patient Figure 8 Patient Pane 1 Patient type parameters Pediatric mode Click if the patient is a child Adult mode Click if the patient is small Click if the patient is medium Cl...

Page 21: ...ult parameter setting is not adapted to your patient type you can manually adapt the parameter setting to the patient type and save this setting as the default setting To save a parameter setting for...

Page 22: ...ogram Resets the preferred program to the manufacturing parameter settings 2 Preview Screen Displays the acquired acquisition frames 3 Selected Parameter Display Displays the current acquisition param...

Page 23: ...osure 13 Fast scan button Minimizes acquisition time to avoid patient movement 14 3D acquisition option indicates 3D acquisition option selected full upper lower jaw acquisition or focused teeth acqui...

Page 24: ...18 Chapter 3 Imaging Software Overview Note Optional for CS 8100 3D Access and CS 8100SC 3D Access...

Page 25: ...o examine a required block of the upper teeth area of interest To examine a required block of the lower teeth area of interest To enable an acquisition at a higher resolution Select if you want to hav...

Page 26: ...cause the default kV and mA exposure settings are set for the selected patient mode The selected parameters must be based on the age and morphology of the patient Figure 13 Patient Pane Note A selecte...

Page 27: ...ter setting is not adapted to your patient or program type you can manually adapt the parameter setting to the patient or program type and save this setting as the default setting To save a parameter...

Page 28: ...22 Chapter 3 Imaging Software Overview...

Page 29: ...ceed to start the CS Imaging Software Increasing the operating life of the X ray tube To increase the operating life of the X ray tube follow these steps 1 In the Panoramic Acquisition Window select t...

Page 30: ...w these steps 1 On your desktop double click to open the CS Imaging Software The Windows Security Alert dialog box displays Select Private networks Public networks and click Allow access 2 Click Quit...

Page 31: ...4 Click the Imaging window to access the Acquisition interface options Click to access the Panoramic Acquisition Interface Click to access the 3D Acquisition Interface The selected acquisition interf...

Page 32: ...26 Chapter 4 Getting Started...

Page 33: ...Panoramic Acquisition Interface Preparing the Unit and Setting the Acquisition Parameters To set the acquisition parameters follow these steps 1 In the Panoramic Acquisition Interface click the Patie...

Page 34: ...ing if you want to save the new parameter settings as future default settings 4 Position and fix the panoramic chin rest A Fit the bite block to the panoramic chin rest Ba Make sure that you click it...

Page 35: ...barrier as shown in C and D If you are using the Frankfort guide bite block for panoramic use the same single use hygienic sheaths as for use with 3D bite blocks C D E Make sure that you cover the bi...

Page 36: ...s shown For a more accurate Camper plane alignment you can Use the panoramic chin rest A Fit the standard bite block or edentulous bite stick to the panoramic chin rest B A B Important Do not use the...

Page 37: ...djustment knob to open the temple supports laterally 4 Ask the patient to enter the unit On the Positioning panel press and hold to raise the chin rest to the height of the patient 5 Ask the patient t...

Page 38: ...panoramic ask the patient to bite on the bite block B as shown Note By correctly positioning the patient you can reduce the spinal column shadow that is transferred to the acquired image Important Yo...

Page 39: ...g indicators as a visual aid to adjust the up or down inclination of the patient head D1 to obtain a Frankfort plane horizontal alignment 9 Use both hands to align the vertical position of the head D...

Page 40: ...he exposure button until the end of the acquisition is indicated by the on screen Release Handswitch message When the acquisition ends the Acquisition interface disappears and the acquired image is au...

Page 41: ...Unit and Setting the Acquisition Parameters To set the acquisition parameters follow these steps 1 In the Panoramic Acquisition interface click the Patient button to access the Patient pane Select th...

Page 42: ...priate parameters Click and select Memorize Anatomy setting to save the new parameter settings as future default settings 6 On the head and chin rest remove or place and cover it with a hygienic barri...

Page 43: ...ent knob to close the temple supports Stand up straight do not position the feet slightly forward Grip both the lower handles of the head and chin rest Relax and lower the shoulders for full motion of...

Page 44: ...ning sound is heard indicating X ray emission Press and hold the exposure button until the end of the acquisition is indicated by the on screen Release Handswitch message When the acquisition ends the...

Page 45: ...e head and chin rest X ray Dose Emission Information Compliance with EURATOM 97 43 Directive You can right click on each image to display the estimated emitted dose received by the patient You can use...

Page 46: ...40 Chapter 5 Acquiring Panoramic Images...

Page 47: ...terface Preparing the Unit and Setting the Acquisition Parameters To set the acquisition parameters follow these steps 1 In the 3D Acquisition Interface click the Patient button to access the Patient...

Page 48: ...A and insert the 3D bite block B as shown Make sure that you click the bite block well into place See the table below for information on the recommended Fields of View FoV and Acquisition types for a...

Page 49: ...r backing from the hygienic barrier C and cover the bite block with the hygienic barrier D Make sure that you cover the bite block completely with the hygienic barrier E E C D Important You must cover...

Page 50: ...ad apron Ensure that the apron lays flat across the patient s shoulders 3 Turn the adjustment knob to open the temple supports laterally 4 Ask the patient to enter the unit On the Positioning panel pr...

Page 51: ...align the vertical position of the head D Ensure that the patient occlusion is fully horizontal camper plane D1 9 Ask the patient to Close the eyes E Swallow Remain still Breathe through the nose Pla...

Page 52: ...ed by the on screen Release Handswitch message The turns yellow and a warning sound is heard indicating X ray emission When the acquisition ends the Acquisition Interface disappears and the volume is...

Page 53: ...Accessed the 3D Acquisition Interface Preparing the Unit and Setting the Acquisition Parameters To set the acquisition parameters follow these steps 1 In the 3D Acquisition interface click the Patient...

Page 54: ...te block use for each Acquisition type and Field of View FoV If you are using a standard 3D bite block or a 3D molar bite block fit the bite block A to the support as shown 3D focused teeth acquisitio...

Page 55: ...mouth To acquire an image of the right side 3rd molar insert the offset bite block so that the part indicated MR will be placed in the corresponding position inside the patient s mouth 6 Remove the p...

Page 56: ...adjustment knob to open the temple supports laterally as shown If you are using the 3rd Molar offset 3D bite block optional Remove the standard temple support from the right or left head clamp depend...

Page 57: ...the chin rest to the height of the patient 5 Ask the patient to 6 To position the patient Stand up straight do not position the feet slightly forward Grip both the lower handles of the head and chin r...

Page 58: ...e block To acquire an image of the right 3rd molar ask the patient to bite on the side of the bite block indicated MR as shown A To acquire an image of the left 3rd molar ask the patient to bite on th...

Page 59: ...aking sure that the patient places the upper incisors before the incisive stopper B 7 Turn the adjustment knob to close the temple supports C 8 Use both hands to align the vertical position of the hea...

Page 60: ...set under reconstruction During reconstruction release the patient and remove the hygiene barrier from the bite block 3 Open and review the volume from you imaging software X ray Dose Emission Informa...

Page 61: ...Firmware versions 2 Animated display screen Guides you on the acquisition positioning modes 3 Selected Parameter Display Displays the current acquisition mode and parameter settings 4 System Status S...

Page 62: ...ns Provides the various acquisition scenarios 13 Acquisition Export Mode Provides one of the following acquisition export modes to view the acquired images CS Restore CS Model Appliance 14 Acquisition...

Page 63: ...Family SM842 Ed07 57 CS Restore Impression Material Pane Overview The CS Restore impression material program pane becomes active when you Click Select Click for the triple tray impression mode Click...

Page 64: ...mode Click for the full jaw mode with wax bite Click for the lower jaw mode Click for the upper jaw mode Click for the lower and upper jaw in occlusion mode Important To have a true occlusion alignme...

Page 65: ...2 Ed07 59 CS Model Impression Material Pane Overview The CS Model impression material program pane becomes active when you Click Select Click for the full jaw mode Click for the full jaw mode with wax...

Page 66: ...0 Chapter 7 3D Object Acquisition Appliance Radiological Guide Pane Overview The radiological guide program pane becomes active when you click Click for the lower jaw mode Click for the upper jaw mode...

Page 67: ...SM842 Ed07 61 3D Object Acquisition Accessories Table 2 3D Object Acquisition Accessories Item Description Quantity 3D dental impression holder support 1 3D dental impression holder 1 Black foam A Ce...

Page 68: ...rt A in the chin rest base 2 Place the dental impression holder B onto the 3D bite block support A 3 Place the black foam C on the dental impression holder B 4 Click the Program button to access the P...

Page 69: ...quisition interface when is green press and hold the exposure button of the X ray remote control until the end of acquisition The indicator in the acquisition interface turns yellow and a warning soun...

Page 70: ...apter 7 3D Object Acquisition 5 Once all items have been scanned the Acquisition interface closes Wait for the 3D object reconstruction The completely reconstructed object will display in the image br...

Page 71: ...ion interface turns yellow and a warning sound is emitted to indicate that X ray emission is in progress The acquired image appears in the Preview Screen 5 Once the triple tray impression has been sca...

Page 72: ...66 Chapter 7 3D Object Acquisition The completely reconstructed object will display in the image browser...

Page 73: ...g window Click then to access the Acquisition Interface Preparing the Unit and Setting the Acquisition Options To prepare the unit for acquisition follow these steps 1 Insert the 3D bite block support...

Page 74: ...the acquisition interface turns yellow and a warning sound is emitted to indicate that X ray emission is in progress The acquired image appears in the Preview Screen 5 Once all items have been scanne...

Page 75: ...arning sound is emitted to indicate that X ray emission is in progress The acquired image appears in the Preview Screen 5 Once the triple tray impression has been scanned the Acquisition interface clo...

Page 76: ...sert the 3D bite block support A in the chin rest base 2 Place the dental impression holder B onto the 3D bite block support A 3 Place the black foam C on the dental impression holder B 4 Click the Pr...

Page 77: ...s yellow and a warning sound is emitted to indicate that X ray emission is in progress The acquired image appears in the Preview Screen If you select the option that requires the acquisition of the wa...

Page 78: ...il the end of acquisition The indicator in the acquisition interface turns yellow and a warning sound is emitted to indicate that X ray emission is in progress The acquired image appears in the Previe...

Page 79: ...on follow these steps 1 Insert the 3D bite block support A into the chin rest base 2 Place the dental impression holder B on the 3D bite block support A 3 Place the black foam C on the dental impressi...

Page 80: ...e control until the end of acquisition The indicator in the acquisition interface turns yellow and a warning sound is emitted to indicate that X ray emission is in progress The acquired image appears...

Page 81: ...remote control until the end of acquisition The indicator in the acquisition interface turns yellow and a warning sound is emitted to indicate that X ray emission is in progress The acquired image ap...

Page 82: ...t the patient record Access the imaging window Click to access the 3D Acquisition Interface Preparing the Unit and Setting the Acquisition Parameters To prepare the unit for acquisition follow these s...

Page 83: ...nal for CS 8100 3D Access and CS 8100SC 3D Access In Canada the full upper and lower jaw acquisition is limited to 8x8 FOV instead of 8x9 FOV Full upper and lower jaw acquisition Recommended Fields of...

Page 84: ...side Position the feet slightly forward Relax and lower the shoulders to allow for the full motion of the unit rotative arm 4 Place the radiological guide A and the bite index B in the patient s mouth...

Page 85: ...old the exposure button of the X ray remote control until the end of acquisition The indicator in the acquisition interface turns yellow and a warning sound is emitted to indicate that X ray emission...

Page 86: ...epare the unit for acquisition follow these steps 1 Insert the 3D bite block support A in the chin rest base 2 Place the dental impression holder B onto the 3D bite block support A 3 Place the black f...

Page 87: ...w and a warning sound is emitted to indicate that X ray emission is in progress The acquired image appears in the Preview Screen 3 Once all items have been scanned the Acquisition interface closes Wai...

Page 88: ...rface Preparing the Unit and Setting the Acquisition Options of the NobelGuide or NobelClinician Calibration Phantom To prepare the unit for acquisition follow these steps 1 Insert the 3D bite block s...

Page 89: ...Program pane 6 Click to access the Radiological Guide acquisition mode interface appears in the Acquisition Export Mode field 7 Select either one of the appliance acquisition options Important The ani...

Page 90: ...ace turns yellow and a warning sound is emitted to indicate that X ray emission is in progress The acquired image appears in the Preview Screen 3 The animated display in the Acquisition interface chan...

Page 91: ...t and dry cloth Annual maintenance We recommend a general inspection of the unit carried out by an authorized service technician Controlling the Image Quality To maintain optimum image quality the ima...

Page 92: ...Chapter 8 Maintenance 1 On your desktop double click The CS 8100 3D Technician Tools window is displayed 2 In the left pane double click Image Quality Tool to launch the Image Quality Control softwar...

Page 93: ...or CS 8100 3D Family SM842 Ed07 87 3 Click Next to access the Image Quality Control window 4 Click test options to select the desired test or to modify default test parameters 5 Click Start a new test...

Page 94: ...88 Chapter 8 Maintenance...

Page 95: ...until the end of the acquisition cycle Message Description Action X ray tube cooling Cooling in progress Wait until the Generator Cooling Indicator on the Acquisition interface reaches zero Thermal s...

Page 96: ...90 Chapter 9 Troubleshooting...

Page 97: ...ance Authorized Representatives European Authorized Representative Carestream Health France 1 rue Galil e 93192 Noisy Le Grand Cedex France Representante no Brasil Carestream do Brasil Com rcio e Serv...

Page 98: ...SM842 Ed07 2017 06 Carestream Dental A Division of Carestream Health Inc 150 Verona St Rochester NY 14608 USA Carestream Health Inc 2017 For more information visit www carestreamdental com...

Reviews: