Panoramic and 3D Modality User guide for CS 8100 3D Family (SM842) Ed07
63
Launching the Acquisition for the Impression Material
To launch the acquisition, follow these steps:
1.
Select either the lower jaw or upper jaw acquisition mode.
2.
Follow the on-screen animated instruction on how to correctly position the impression at the
center of the black foam.
3.
Leave the X-ray room and close the door.
4.
On the
Acquisition
interface, when
is green, press and hold the exposure button of the X-ray
remote control until the end of acquisition.
The
indicator in the acquisition interface turns yellow and a warning sound is emitted to
indicate that X-ray emission is in progress.
The acquired image appears in the
Preview Screen
.
WARNING:
•
ALWAYS use a metal-free tray for the Impression
Material.
•
Make sure that the Impression Material and the plastic
tray are dry.
Important: If you have a problem that requires you to stop the
acquisition, release the exposure button of the remote control or
press the red emergency stop button.