-8-
Instruction Manual
Specifications
e. Operator presses the "OFF" switch, and allows the machine to sit for a minimum of 6 hours before using
again.
NOTE
– A sanitize holding period of six hours or more will expose the machine’s fluid pathways to Proportioned
RENALIN 100 Solution for a time consistent with the disinfection time indicated in the RENALIN
®
100 Cold
Sterilant labeling. If the "START PROCESS" switch is pressed after entering "PROGRAM STEP" 83, total
sanitize cycle time will be approximately seventeen (17) minutes which includes a 10 minute hold time. This will
sanitize the machine fluid pathways. This cycle will also allow the purging of the RENATRON
®
II 100 Series
hydraulic pathways for shipping purposes. Use the methods described in c - d above for a routine sanitization
procedure.
f. Operator wipes clean the external machine surfaces with a lint free disposable towel saturated with fresh
1% RENALIN 100 solution or undiluted ACTRIL
®
Cold Sterilant.
g. The RENATRON II 100 Series System is now ready for reprocessing of dialyzers.
A.
Materials of Construction
: Acrylic, PVC, ABS, Polypropylene, Delrin
®
, Ethylene Propylene, Viton
®
Rubber,
Stainless Steel.
B.
Electrical Leaka
ge Current
(
risk current
): Each RENATRON II 100 Series is allowed a leakage current less
than 100 micro amps RMS MAX at 115 VAC per UL standard 61010-1.
C.
Electrical Requirements
:
1. RENATRON II 100 Series station:
Voltage:
Selectable 100, 120, 230, 240 VAC 50/60 Hz single phase
Fuses:
100-120 VAC, one 4A 3AG (1.25 inches x 0.25 inches)
220-240 VAC, two T2A (5mm x 20mm)
Equipment Classifications:
Class 1
Electrical Leakage Current
(risk current):
From cabinet to ground: <100 µA
From ground to outlet: < 100 µA
Power Consumption:
460 VA maximum
Mode of Operation:
Continuous operation
Transient Overvoltage
Category 2
Pollution Degree
2
2. Interface Module:
Voltage:
Selectable 115, 230 VAC 50/60 Hz single phase
Fuses:
175 ma, 250V Time Delay Fuse
Equipment Classification:
Class 1
Electrical Leakage Current
(risk current):
< 500 µA
Power Consumption:
20 Va maximum
Mode of Operation:
Continuous operation
3. Label Printer:
100-240 VAC., 50/60 HZ., 1.5 AMP, 39.1 Watt Max.
4. Report Printer:
100-240 VAC., 50-60 HZ., 30 Watt Max.
5. Monitor-CRT:
100-120 VAC., 50/60 HZ., 0.8 AMP, 100 Watt Max.
6. Monitor-Flat Screen:
100-240 VAC., 50-60 HZ., 30 Watts Max.
7. Computer:
100-120 VAC., 50/60 HZ., 230 Watt Max.