-30-
Instruction Manual
D.
Culture Sample Collection RENATRON
®
stations - RS-8335
Water culture samples can be easily collected from the RENATRON station by first running the RENATRON Water
Culture Sample Cycle.
1. Prepare the RENATRON for sample collection
a. Ensure that the RENATRON was sanitized with the calibration cell in place, allowed to dwell for at least
6 hours, and was undisturbed since the initiation of the Sanitize Cycle.
2. Initiate the Water Culture Sample Cycle
a. Press the
"ON"
switch.
b. Press the
"MUTE ALARM"
and
"RESET"
switches to change to the "CH" mode.
c. Press the following switches in the following order:
"MUTE ALARM"
"RESET"
"DISPLAY"
d. The Volume Display window will read "CUL" and the Program Step display window will read "CC".
e. Press the
"START PROCESS"
switch
f. At the end of the program step
9
, the
"PROCESS COMPLETE"
message will illuminate and the Program
Step display window will read "SA".
3. Sample Collection
a. Using aseptic technique, collect a sample from the calibration cell for a culture of water from the
RENATRON station.
b. After the sample has been collected, press the
"RESET"
switch. The RENATRON II 100 Series Station
is now ready for calibration verification and then for reprocessing dialyzers.
E.
Storage When Not In Use
The RENATRON II 100 Series station should be stored in a protected location. Avoid high-traffic areas where the
machine might get damaged and avoid high humidity.
The RENATRON II 100 Series station is a high-quality, precision device and should be treated as such. The station
has been designed to withstand the rigors of normal use and, with reasonable care and maintenance, should provide
many hours of trouble-free operation.
MEDIVATORS maintains a special repair and calibration facility to provide its customers with expert service for its products.
If your RENATRON II 100 Series station requires servicing, we recommend that you contact our Renal Technical
Support. MEDIVATORS conducts periodic service/maintenance seminars for the RENATRON II 100 Series. Contact
our Renal Technical Support or your sales representative for information regarding these classes.