Instruction Manual
3. "TANK VOLUME" error message appears when the internal mixing/measuring tank should be empty but is
not or when an improper action has occurred within the tank.
4. "CLEAN" message appears during the dialyzer cleaning steps.
5. "ADD CHEMICAL" error message appears when machine detects that an insufficient amount of RENALIN
®
100 Cold Sterilant has been drawn into the mixing/measuring tank.
6. "TEST" message appears during the dialyzer testing steps.
7. "VOLUME FAIL" error message appears when measured dialyzer blood compartment (priming) volume is
less than the blood compartment reference volume.
8. "PRESSURE FAIL" error message appears when a dialyzer fiber leak or compromised dialyzer header
integrity has been detected by the machine.
9. "DISINFECT" message appears during the dialyzer disinfection steps.
10. "ALARM" error message flashes continuously while the machine is in alarm mode.
11. "PROCESS COMPLETE" message appears when dialyzer reprocessing cycle is complete.
12. "INTERLOCK" error message appears when the "START SANITIZE" switch has been pressed and the
sanitize interlock is not engaged.
13. "SANITIZE COMPLETE" message appears when the machine sanitization cycle is complete.
D.
Interface Module Indicators
1. "POWER" is the light emitting diode (LED) on the front panel. When illuminated, it indicates electrical power
is switched on.
2. "SCANNER ON LINE" is the light emitting diode (LED) on the front panel. When illuminated, it indicates
bar code scanner can be used to input data. Bar code scanner is disabled at all other times.
NOTE – Separate manuals are provided with the computer system. Refer to the respective manuals for detailed
information about the computer, monitor, and printer controls and indicators.
IMPORTANT – TO ISOLATE THE STATION FROM THE ELECTRICAL SUPPLY YOU MUST DISCONNECT THE
CORD FROM THE POWER ENTRY MODULE LOCATED IN THE LOWER RIGHT (WHEN VIEWED FROM THE
FRONT) SIDE OF THE UNIT BACKSIDE. THE UNIT SHOULD BE SITUATED TO ALLOW AS MUCH ROOM ON
THE RIGHT SIDE AS POSSIBLE, ALLOWING EASY ACCESS.
CAUTION
– TOTAL SYSTEM CHASSIS RISK CURRENT SHOULD NOT EXCEED 100 MICRO AMPS.
ATTENTION
– LE COURANT DE RISQUE TOTAL ADMISSABLE AU CHASSIS NE DOIT PAS DEPASSER
100 MICRO AMPS.
NOTE – Report any damage promptly to the shipping company and make claim for compensation. Damage
incurred during shipment is not covered by the machine warranty. Please save the shipping container for use
in returning machine if repairs are necessary.
NOTE
– All plumbing (hoses and lines) are cut to length and calibrated at the factory and must not be modified from
the original lengths or sizes. Only factory authorized service personnel should repair or replace these items.
CAUTION
– PRIOR TO CONNECTING, ALL WIRING MUST BE CHECKED TO ENSURE PROPER POLARITY
AND ALL GROUNDS MUST BE CHECKED TO ENSURE CONTINUITY.
CAUTION
– WHEN THE WATER SUPPLY FEED LINE IS CONNECTED TO THE RENATRON
II 100 SERIES, IT MUST BE EQUIPPED WITH AN "ON-OFF" VALVE THAT WILL PERMIT
SHUTTING OFF THE FLOW OF WATER TO THE RENATRON
II 100 SERIES WHEN NOT IN USE.
CAUTION
– PROPER WATER SYSTEM OPERATION MUST BE VERIFIED. SEE SPECIFICATION SECTION
FOR WATER REQUIREMENTS.
NOTE
– THE RENALIN
100 COLD STERILANT CONTAINER IS TRANSLUCENT TO ALLOW THE
USER TO VIEW THE LEVEL OF SOLUTION. ALWAYS USE THE RENALIN 100 COLD STERILANT
CONTAINER AS A SUPPLY TANK FOR THE RENATRON II 100 SERIES. DO NOT USE ANY OTHER
TYPE TANK OR CONTAINER.
RENATRON
®
II
100 Installation
Instructions